Part 1 - Introduction
Max. Sensor-to-Analyzer
100 feet (30.5 meters), with junction box
Distance
Instrument Power
Two generic AA batteries, low battery indication at 1.60V
DC.
Outputs
Two 0-2.5 VDC isolated outputs are provided
on for connection to data recorders, etc.
Battery Life
Approximately 240 hours of operational use on a set of
batteries (without backlight).
1.4
Q45D Performance Specifications
Accuracy
0.2% of span or better
Repeatability
0.05% of span or better
Sensitivity
0.05% of span
Non-linearity
0.1% of span
Stability
0.05% of span per 24 hours, non-cumulative
Temperature Drift
Span or zero, 0.01% of span/ºC
Warm-up Time
3 seconds to rated performance
Instrument Response
6 seconds to 90% of step input, in a given
Time