ATI
Model F12/D Gas Transmitter
Part 3
– Electrical Connections
17
O&M Manual (Rev-H)
3.3
Terminal Board
– Loop Power Connections
The Loop Powered transmitter model is powered in using terminals 5 and 6
commonly referred to as “two-
wire mode”.
Loop Powered models with the heated sensor housing and/or MODBUS communications option require an
additional 24 VDC supply to power them. This auxiliary power is connected to terminals 1 and 2.
Note:
Connections to the communications on the DC Powered models are made on the power supply
board.
Power Terminals
1
Auxiliary Power +
2
Auxiliary Power -
3
Comm. Tx
4
Comm Rx
5
Loop +
6
Loop -
7
Comm Common
8
EARTH GROUND
Figure 14. Power and communication terminals.