Manual, F/T Sensor, Data Acquisition (DAQ) Systems
Document #9620-05-DAQ.indd-20
Pinnacle Park • 1031 Goodworth Drive • Apex, NC 27539 • Tel: 919.772.0115 • Fax: 919.772.8259 •
26
Figure 4.3—Unterminated Cable Wiring (9105-C-PS-U-x) and USB DAQ Device System Installation
USB DAQ Device [1]
[3b] Insert unterminated Cable Ends Into
Appropriate Screw Terminal Location
and Tighten Set Screw
[2,3,4,6] DAQ PS or
IFPS Box (9105-PS-1 Shown)
[4] 12 Volt Wall Mount Power Supply,
2A,24 W (9105-PS-KTPS24-12020MP)
[3b] Protect Exposed Unused Wire Ends
and Tie Wires Back to Cable Jacket
[3b] Twist Shielding Together
and Insert Into DGND
[3b] If 12 Volt Wall Mount Power Supply
is Being Used, do not Connect the
Red Wire to DGND or the
Red/White Wire to +5V
Protect Wire Ends and Tie
Wires Back to Cable Jacket
[1] USB Cable
Note: [#] Indicates the Step Number.
NOTICE:
If you are not using a National Instruments DAQ board with mass termination,
you must provide your own connector at that end of the cable. See
for connection information.
Figure 4.4—Unterminated Cable Wiring (9105-C-PS-U-x) for a NI 32 Screw Terminal USB DAQ Device
(SG1 Output) Yellow to AI1
Protect Unused Wire Ends
and Tie Back to Cable Jacket
Twisted Shielding to DGND
(SG0 Output) Brown to AI0
(SG0 Ref) Brown/White to AI8
(SG1 Ref) Yellow/White to AI9
Green/White to AI10 (SG2 Ref)
(SG2 Output) Green to AI2
Blue to AI3 (SG3 Output)
Blue/White to AI11 (SG3 Ref)
Violet to AI4 (SG4 Output)
Violet/White to AI12 (SG4 Ref)
Grey to AI5 (SG5 Output)
Grey/White to AI13 (SG5 Ref)
Black to AIGND (AIGnd)