Integrated signalling equipment
(acoustic)
D=FLED.A: 70 dB 1 m distance
Protection class
III
Dimension
See figure 4
Cable glands
• Cable diameter range
• External thread
See section 10
M20 x 1.5
Acceptable cables
0.5 – 2.5 mm
2
(cable lugs 2 x 1 mm
2
or 2 x 0.75 mm
2
should be used for double wires)
Enclosure material
• Aluminium spray epoxy
• Aluminium creodur epoxy
• SS316L
Measuring head material
SS316L
Weight
3.5 kg
Mandatory periodic inspection
Every 12 months (Calibration Certificate validity) – time can be shortened
due to difficult working conditions
Lifetime of consumables
See table 12
Mounting
• To the supporting structure, 2 screw holes 4 mm, hole spacing 127 mm
• We recommend using mounting brackets – see table 13
Table 10: Technical specification
In the table below shows the gas detector current consumption depending on the device configuration.
Product code
Current consumption [mA]
15 V
24 V
PW-044
SG4
EC
0
FL, FL.M, HL
20
20
LCD
45
35
FLED, FLED.A
125
75
PEL
0
100
70
LCD
115
80
FLED, FLED.A
205
125
IR
0
FL, FL.M, HL
80
55
FH, FH.M, HH
155
95
LCD
FL, FL.M, HL
95
65
FH, FH.M, HH
170
105
FLED, FLED.A
FL, FL.M, HL
185
110
FH, FH.M, HH
260
150
PID
0
FH, FH.M, HH
120
75
LCD
145
90
FLED, FLED.A
225
130
Table 11: Power consumption
p
. 25|28
User Manual: POD-061-ENG R03