RU device is turned off.
The incorrect wiring between the RU and the IDA8.
Indication:
LED Indicator: When this error is detected, the following LEDs will light up:
o
G. Fault LED on the front panel of the IDA8C/S.
Log:
The text messages display at the touch screen of the IDA8C:
"Redundancy Error"
"Backup Device Offline." or
"Operation Mode Fail."
A text message displays at the window [Machine Fault Record] in Ateis Studio:
A buzzing sound outputs to the monitoring speaker on the front panel of the IDA8C.
Troubleshooting:
The only power supply for RU is from IDA8s. In case the 24VDC output connector and wire are not
working right, the IDA8 can detect it and show the fault list on the touch screen as following:
VS_D12V_RL Error
-12V Error
12V Error
If anyone of them shows, please check the connections or replace the device.
In case of a damaged relay of RU which cannot switch the system, the Primary IDA8 would
become the main system.
Once it's recovered, the log will show:
"RU OK: Backup Device Online."
"RU OK: Operation Mode Normal."
3.1.9.1.13 URGP16I16O Fault
There is an monitoring error in URGP16I160 device due to the reasons of out of tolerance.
Indication:
LED Indicator: When this error is detected, the following LEDs will light up:
o
G. Fault LED on the front panel of the IDA8C/S.
Log:
A text message displays on the touch screen of IDA8C:
"URGP16I160 Fault"
The text messages display at the window [Machine Fault Record] in Ateis Studio:
o
Port: Indicate which port in your device has fault.
o
ID: Indicate which customized name triggers fault.
o
CHn: Indicate which channel is cut off.