Atal TE232-105D Скачать руководство пользователя страница 1

 

 
 
 
 

Programmable indoor transmitter of temperature 

TE232-105D, TE485-105D 

 
 

Programmable indoor transmitter of temperature, 

relative humidity and other derived humidity values 

TR232-105D, TR485-105D 

 
 
 

Programmable indoor transmitter of atmospheric 

pressure, temperature, relative humidity and other 

derived humidity values 

TRP232-105D, TRP485-105D 

 
 

 

with RS232 / RS485 serial output

 

 

 

 
 
 
 
 

Instruction Manual 

 
 
 

Содержание TE232-105D

Страница 1: ...ansmitter of temperature relative humidity and other derived humidity values TR232 105D TR485 105D Programmable indoor transmitter of atmospheric pressure temperature relative humidity and other derived humidity values TRP232 105D TRP485 105D with RS232 RS485 serial output Instruction Manual ...

Страница 2: ... use different communication protocol then Modbus RTU it is necessary preset them throw user s software see latter Temperature relative humidity eventually pressure sensors are non removable device parts pressure sensor is placed inside the electronic case Measured and computed values are optionally shown on dual line LCD display If there are two values displayed at one LCD line they are periodica...

Страница 3: ...wer terminal to DTR into device terminal only and then do NOT interconnect DTR and RTS with master device Devices with RS232 interface Termina l Connect to GND Shielding power source ground Tx Output data for master device transmit Rx Input data from master device receive RTS Connect with RTS signal to master device eventually o not connect DTR Connect with DTR signal to master device or connect p...

Страница 4: ...mal length is restricted to 15 m and allow connect only one device Electrical system wiring may do only worker with required qualification by rules in operation Dimensions TE485 105D TR485 105D TRP485 105D RS485 TE232 105D TR232 105D TRP232 105D RS232 4 ...

Страница 5: ...f there is necessary demount the device insert flat bladed screwdriver max 3 5 mm wide from top side into middle device s air hole There is fastening member placed insert screwdriver beyond the fastening member about 2 cm deep number c at the picture Then slightly move screwdriver in arrow direction d as shown at the picture It unlock fastening member and the device is partially opened Remove the ...

Страница 6: ...should be opened Actual adjusted address of the device is displayed on LCD display at decimal base Next press of button exits info mode and actual measured values are displayed Note No measurement and communication is possible during info mode If device stays in info mode for longer than 15 s device automatically returns to measuring cycle Setting of pressure correction for altitude For devices TR...

Страница 7: ... input gates Read Input Registers 16 0x10 Setting of more 16 bit registers Write Multiple Registers Jumper and button Jumper and button are located next to connection terminals If communication protocol Modbus is selected the function of jumper and button is as follows Jumper opened device memory is protected from writing from device side it is only enabled to read measured value writing to memory...

Страница 8: ...T7410 with atmospheric pressure measurement Variable Unit Address hex X Address dec X Format Size Status hPa Int 10 PSI Int 1000 inHg Int 100 mBar Int 10 oz in2 Int 10 mmHg Int 10 inH2O Int 10 Atmospheric pressure kPa 0x0034 52 Int 100 BIN16 R Explanation depends on device setting by User s software Int 10 register is in format integer 10 R register is designed only for reading W register is desig...

Страница 9: ...dress to 00h it will communicate in the same communication speed as before closing jumper and will communicate without check sum After jumper is opened setting of address and check sum is reset in accordance with values stored in the device Communication speed and check sum are possible to change only if jumper is closed Jumper closed and button pressed for longer than six seconds causes restoring...

Страница 10: ... of the value is not possible error during measurement of temperature or relative humidity or value is over range Error 2 There is a reading Err2 on LCD display Measured or calculated value is below lower limit of allowed full scale range Value read from the device is 999 9 This state appears in case of Measured temperature is lower than approximately 210 C i e low resistance of temperature sensor...

Страница 11: ...rror state of value If selected pressure unit is mBar or oz in2 or mmHg or inH2O or kPa there is shown only value number without corresponding pressure unit C F DP Reading next to this symbol is calculated dew point temperature or error state of value g m3 Reading next to this symbol is calculated absolute humidity or error state of value g kg Reading next to this symbol is calculated specific hum...

Страница 12: ... 0 8 0 5 0 20 The value computed from ambient temperature and relative humidity TR232 105D TR485 105D TRP232 105D and TRP485 105D devices only Display resolution 0 1 C You can choice one of the next value Dew point temperature Accuracy 1 6 C at ambient temperature T 25 C and RH 30 Range 60 to 80 C Absolute humidity Accuracy 3g m3 at ambient temperature T 40 C and RH 95 Range 0 to 400 g m3 Specific...

Страница 13: ...tromagnetic compatibility complies EN 61326 1 Storing conditions temperature 30 to 80 C humidity 0 to 100 RH without condensation Not allowed manipulations It is not allowed to operate the device under other than specified conditions in technical parameters Devices are not designed for locations with chemically aggressive environment Temperature and humidity sensors must not be exposed to direct c...

Страница 14: ...essory for connection of transmitter transmitters with RS485 interface to the PC via serial port RS232 Connect connector marked RS232 directly to the PC connect power to connector marked RS485 Power voltage 6V DC from an external acdc adapter connect to pin 9 0V connect to pin 5 Also mutually connect pin 2 and pin 7 Link RS485 is connected across pin 3 A and pin 4 B 14 ...

Страница 15: ...Notes 15 ...