P-SOP0804-3262-E02 

1/5 

Standard Operating 

Procedure    

SOP

 

 

 

Automatic digital 

Refractometer 

RX-7000

α

(alpha) 

 

 

 

Preface 

 

To ensure proper care operation, many years accurate measurement, the Standard Operating 

Procedure (SOP) document for the automatic  digital refractometer,RX-

7000α

,  is provided 

herewith.  The following instructions should be used as an example.  Based on the steps 

provided,

 

write  up your company’s own original SOP for the operation of the automatic 

digital refractometer, RX-

7000α.

 

 

 

The standard operation procedure document for the automatic 

digital refractometer, RX-

7000α

 

 

 

1. Operation Manual 

 

Plug the AC power supply cable into an AC100 to 240V power outlet.   If the  optional 

digital printer for the RX-7000

α

  is connected, turn on the power supply switch of the 

printer first, and then, turn on the power supply switch of the RX-

7000α

.   

 

1)

 

After checking  the initial screen (opening message) and also checking the present 

temperature in the measurement display screen,  perform  zero  adjustment with 

distilled water. 

2)

 

Drop 0.1ml or more distilled water onto the prism surface and close the cover lid, 

then, press the SW1 (ZERO) key.  Screen displays “WAITING…”. After the current 

temperature(PRESENT) reaches the targeted temperature (TARGET), the screen 

displays ”READING…”.    After a few seconds, screen displays “ZERO SET END”.    

Confirm the display of "ZERO SET END". 

Wipe off the distilled water from the prism surface with tissue paper. 

3)

 

Drop  0.1ml or more of sample onto the  prism  surface  and  close the cover lid, then, 

press the START key.  Screen displays “WAITING…”. After the current temperature 

(PRESENT) reaches the targeted temperature (TARGET), the screen displays 

“READING…”.    After a few seconds, the measured value will be indicated. 

4)  Record the  measured value.  If the printer is connected, it will  print the 

measurements taken.  Wipe off the sample from the prism surface with tissue paper. 

5)  Measure the next sample. 

6)

 

After measurement is completed, wipe off the sample from the prism surface and the 

surrounding area with soft tissue paper soaked in water and then wipe dry. 

7)

 

Turn off the power supply switch of the RX-

7000α

.  If the printer is connected, turn off 

the power supply switch of the printer. 

 

*Keep the "instruction manual" of the RX-

7000α in a convenient

 place for easy reference. 

 

 

 

 

 

 

 

 

 

 

Содержание RX-7000a

Страница 1: ...more distilled water onto the prism surface and close the cover lid then press the SW1 ZERO key Screen displays WAITING After the current temperature PRESENT reaches the targeted temperature TARGET the screen displays READING After a few seconds screen displays ZERO SET END Confirm the display of ZERO SET END Wipe off the distilled water from the prism surface with tissue paper 3 Drop 0 1ml or mor...

Страница 2: ... point is confirmed by the calculation software of the refractometer that is all for calibrating all measuring spans 3 Periodical Inspection Inspect the RX 7000α with standard sucrose solution once a month to see if it indicates the measured value correctly Select a concentration point which you often measure and use this as a reference for inspection Save the checked data and record 3 1 Inspectio...

Страница 3: ...s completed be sure that nothing on the prism surface and surrounding areas are completely wiped off with tissue paper moistened with water then wipe dry with tissue 2 When the body of the instrument is dirty wipe it with dry tissue cloth etc 3 If the digital printer is used the thermo sensitive roll paper is a consumable item 4 Place the refractometer at a designated location Keep it in a place o...

Страница 4: ...ration mistake etc a Sample from the previous measurement may have been left on the prism surface b Distilled water was left on the prism surface c Measurement was executed without closing the cover lid 2 If the operator of the RX 7000α makes any mistakes in operation or inputting of work the RX 7000α displays the corresponding error message a Zero adjustment was performed with a sample b Measurem...

Страница 5: ...ometer RX 7000α alpha 7 Establishment and record of revision When this standard operation procedure document is revised record the revised or changed contents in this sheet and sign approval column Record of revision Approval column Author s signature Approved by Date Establishment ...