External Light Interference (ELI)

The following warning message will be displayed when intense light passes through the prism.

1.Contents

2.Parts 

6. Installation

4. Zero Setting

5. Measurements

7.Error Messages

Digital Immersion Refractometer

Adjust the height of the display unit at the 
clamp and fasten the mounting screw to 
securely fix the shaft in position.

The clamp mount spins free when the 
mounting  screw is loosened, allowing for 
shaft height adjustment at any desired 
angle. Position the display unit upright after 
the height adjustment is complete.

Height-adjusting shaft is fixed 
with the mouting screw.

۔

"LLLL" 

 Measurement Error

There is an insufficient amount of sample on the prism surface

    to perform measurements.

۔

"HHHH" 

Measurement Error

The measured value is out of the measurement range.

۔

Zero Setting Error

There is no or insufficient amount of water on the prism surface 

to perform zero-setting.

Shade the sample stage with your hand and press the START or ZERO key again.

Press the 
START key

Water

Press the 
ZERO key

Hold the ZERO key down 3 seconds

P A N

ZERO

ยܭ

Brix

ZERO

Brix

REFRACT

OMETER

START

PAN-1

Brix 0.0

42.0

Zero-set the unit twice a day.

Blinks three times

Blinks three times

Automatically measures again in 35 seconds

Completion 

If water does not measure 0.0%

Brix

Temperature

Temperature

Temperature

Temperature

Brix

Brix

Brix

Press the 
START key

Brix

࡜ ᗘࢆ஺஫࡟⾲♧

Sample

Brix

Brix

R e p e a t

ͤ

Press the START key to measure again 

before 35 seconds elapse.

Zero-setting is complete if water
measures 0.0%.

Do not dip the instrument in any 
sample that is over 100

Υ

.

Instruction Manual

Fig.A

Only the prism head of the instrument should be submerged into water.

Note regarding water resistance

Thank you for purchasing the Digital Immersion Refractometer, PAN-1DC. Before operating, read this 
instruction manual carefully to understand its contents. Keep the manual with the instrument for future 
reference.

To use the instrument safely, the precautions described in this instruction manual must be observed. 
Failure to comply may result in injury and/or damage to property.

Introduction

Safety Precautions

Use caution as hot samples can cause severe burns. Do not dip the instrument directly into 

a boiling pot to measure.

WARNING

CAUTION

ۍ

 Do not dip the instrument in any sample that is over 100

Υ

.

ۍ

 ATAGO is not liable for any loss and damage caused by the measurement and use of this instrument.

ۍ

 If this instrument is used to measure highly acidic samples, the prism and prism head may be damaged, resulting in 

inaccurate measurements. 

ۍ

 

The prism is made of optical glass. Metal tools and/or implements can damage the prism surface. If the surface of the 

prism is scratched or damaged, inaccurate measurements will occur.

 

ۍ

 

Before use, carefully read the instruction manual and fully understand the function and operation of the instrument. 

ۍ

 Supply the instrument with DC24V only (allowable fluctuation is 

10 %).

ۍ

 Do not use power cables that are damaged or modified.

ۍ

 Do not insert or disconnect the power plug with wet hands.

ۍ

 

Do not leave the instrument in a location exposed to direct sunlight or near a heat source for any extended period of 

time.

?

ۍ

 

Do not change the ambient temperature of the instrument suddenly.

ۍ

 

Do not place the instrument where it will be subject to strong vibrations.

ۍ

 

Do not use the instrument where there is an excessive amount of dust.

ۍ

 

Do not store the instrument in an extremely cool area.

ۍ

 

Do not set or drop heavy objects on the instrument.

The instrument is equipped with the External Light Interference (ELI) 

function to  ensure accurate readings

ELI Function

If the instrument is subjected to intense light, such as direct sunlight or 
artificial lighting, when measuring a sample or zero setting, the ELI 
function will display the [nnnn] (Fig.A) warning message immediately after 
the START or ZERO key is pressed. When this happens, shade the prism 
head and press the START or ZERO key again. If the warning message 
continues to be displayed, place the sample in a non-translucent cup to 
measure.

The instrument comes with the following items:

Blinks three times

*Caution

Press  the ZERO 
key  to  display  the 
temperature first 
before zero-setting.

Press the ZERO key 
between 
measurements to 
display the 
temperature.
*Once the 
temperature is 
displayed, the 
measurement is 
paused. Press the 
ZERO key to resume 
measurement.

Mounting screw

P A N

ZERO

ยܭ

Brix

ZERO

Brix

REFRACT

OMETER

START

PAN-1

Brix 0.0

42.0

P A N

ZERO

ยܭ

Brix

ZERO

Brix

REFRACT

OMETER

START

PAN-1

Brix 0.0

42.0

3. Power and RS-232C output

The power and RS-232C cable connects to the side 
of the instrument, and it provides power (DC24V) to 
the instrument and transmits Brix data via RS-232C.
One multipurpose socket: serves for both power and 
RS-232C output (cable provided in standard 
delivery).
The length of a cable is 2m.

The cable can be ordered to a custom length of up 
to 15m.

1
2
3
4

Red

Black & White

Black

Red & White

DC24V
RS-232C GND
DC24V GND
RS-232C

Pin number

Signal name

Cord color

1

2

3

4

ZERO

Brix

REFR

ACTOMETER

START

PAN-1

Brix 0.0

42.0

1

5

6

9

Z7001 (Clamp filter)

Z7114 (Cable)

20

25

3

Black & White

Red & White

(1)Power

When the DC24V power is supplied, the 
instrument turns on.
The instrument has no power switch.

(2)RS-232C Output

Preparing a PC for Data Transmission

Download a terminal emulator for PC serial 
communication.
Here, the open-source software  "Tera Tarm " 
is used as an example.

Download Tera Term from a website, such as 
the one below:
http://ttssh2.sourceforge.jp/index.html.en/

Start Tera Term. (Fig.3-1)

Select "serial" on the New connection 
dialogbox.
Select the port number.
Click OK.

Check the port number.

"Control Panel" 

 

"System and Security" 

"Device Manager" 

"Port"  

Click Setup, and then Serial port.

Enter the port number selected in step

(Fig.3-2)

Baud rate : 2400bps
Data length : 8bit
Parity : none
Stop bits : 1bit

Click OK.

Click Setup, and then Terminal port.

Enter the settings as shown below and Click 

OK. (Fig.3-3)

Make sure that the "Local echo" is checked.

(3)Transmit

Every time a measurement is taken, a new row 
of data appears in the Tera Term window.

Fig.3-1

Fig.3-2

Fig.3-3

ۍ

 

When measuring hazardous materials, use proper safety procedures, materials, and clothing to avoid personal injury. 

Anyone handling hazardous materials should understand their properties and safety requirements.

ۍ

 

If the instrument is dropped or subjected to a strong impact, contact your supplier for inspection.

ۍ

 

Do not attempt to repair, modify, or disassemble the instrument.

ی

 Digital Immersion Refractometer PAN-1DC (with Clamp) 

͐͐͐

1

ی

 Instruction Manual (this book) 

͐͐͐͐͐͐͐͐͐͐͐͐͐͐͐

1

ی

 Power and RS-232C cable 

͐͐͐͐͐͐͐͐͐͐͐͐͐͐͐͐

1

P A N

ZERO

 ᐃ

Brix

ZERO

Brix

R EFR

AC TO

M ETE

R

START

PAN-1

Brix 0.0

42.0

PAN-1DC
PAN-1DC(M)
PAN-1DC(L)

Cat.No.3606

Cat.No.3607

Cat.No.3608

This part is submerged into 
liquid.

P r i s m

Immersed part

The sensor that detects 
concentrations (Brix %) is 
located.

Press to perform zero-setting.

Press to begin measurement.

  

ZERO Key

START Key

Clamp

This grips the rim of a container.

SUS316L

Silicon
 (O-ring)

ABS

Displays the measured values.

Liquid Crystal Display (LCD)

Battery compartment
PAN-1DC series is not battery 
powered. 
Do not force it to open.

Terminal to connect the 
power and RS-232C cable.

Power and RS-232C terminal

D-SUB 9-pin 
connector solderside

Transmitted data format.

Measurement successful 

 

Temperature

Brix

"LLLL" error 

 

 

Temperature

LLLL

"HHHH" error 

 

 

Temperature

HHHH

External Light Interference (ELI) error 

Temperature

NNNN

Zero-setting successful 

 

ZERO OK

Zero-setting failed   

 

ZERO NG

26.2,0.0
28.5,LLLL
28.6,HHHH
28.9,NNNN
ZERO OK
ZERO NG

Example

Power and 
RS-232C terminal

3606-E03