29
ENGLISH
4.2.7 Hygienic cleaning of the rinsing bottle
1. Remove the lid of the rinsing bottle and
rinse both parts under running water.
2. Let the components dry.
3. Place the lid back on the rinsing bottle.
ASSKEA GmbH recommends replacing the
rinsing bottle every 4 weeks, if it is used and
cleaned hygienically frequently.
4.3 Reuse of the device
The ASSKEA secretion aspirator M30
Plus
is
suitable for reuse. However, prior to passing
the device on to other patients or persons,
it must be processed professionally. Please
send the ASSKEA secretion aspirator M30
Plus
to ASSKEA GmbH or to qualified personnel
authorized by ASSKEA GmbH. Follow the in-
structions in section 6.1!
4.4 Maintenance and service
Risk of performance impairments due
to aging effects
If you notice impaired performance of
the device or other problems, follow
the instructions on troubleshooting
in section 5 and contact the service, if
necessary.
Risk associated with maintenance
and service during operation
Maintenance and service must not be
performed while the device is in use!
The device must be switched off befo-
re maintenance and service work are
performed.
Risk associated with performance of
maintenance and service by unautho-
rized persons
Maintenance and service may be per-
formed only by specialized personnel
authorized by ASSKEA GmbH or by
ASSKEA GmbH itself.
Risk associated with performance of
maintenance and service without suf-
ficient documents
Maintenance and service must be
performed taking into account the
relevant service instructions for the
ASSKEA secretion aspirator M30
Plus
.
ASSKEA GmbH will also provide addi-
tional documents on request, if these
can assist the service staff.
WARNING: Modification of the ME de-
vice is not permitted!
The ASSKEA secretion aspirator M30
Plus
is
maintenance-free, if used according to the
instruction for use with the exception of com-
ponents with limited shelf-life.
Perform a visual and functional check prior to
each use. Make sure to include the accesso-
ries for the device.
Opening the device and repairs must be
performed only by ASSKEA GmbH or
by specialized personnel authorized by
ASSKEA GmbH in compliance with the ser-
vice documentation specified by the manu-
facturer as well as with technical and hygie-
nic precautionary measures.
The device may be sent back for repair to ASS-
KEA GmbH directly or via the specialty dealer
from whom you purchased the device.
Clean and disinfect all accessories prior to re-
turning the device. The device itself must be
treated with a surface disinfectant (see sec-
Содержание M30 Plus
Страница 1: ...M30Plus Gebrauchsanweisung Instruction for Use ...
Страница 43: ...DEUTSCH ...
Страница 83: ......