Page 20 of 38
Symbol
Description
Meaning of display
display illuminates
continuously green
Power supply unit is connected, voltage available.
display blinks red
Incorrect power supply unit is connected, no charging of
battery possible!
button illuminates
continuously green
Device is switched on and ready for use.
button illuminates
continuously orange
Device still runs for 30 sec. before it switches off
automatically. Afterwards the button will still be
illuminated for 10 seconds.
button illuminates
continuously red
Internal error, operation is not possible.
Please contact your service partner!
3.3.2
Setting the vacuum
1.
Switch on the device by pressing the on/off button (fig. 11(F)).
2.
Select the desired vacuum range by pressing the corresponding button (`Low´,
`Medium´ or `High´) on the control panel (fig. 11(A, B, C)).
As described in section 3.3.1, the vacuum setting buttons correspond to certain
vacuum ranges.
3.
You can perform fine adjustments of the set vacuum via the additional air opening on the
fingertip (fig. 12).
Fig. 12
4.
Switch off the device by holding the on/off button (fig. 11(F)) pressed for 1-2 seconds.
3.3.3
Aspiration
1.
Perform the aspiration procedure as you were instructed by specialized personnel.
2.
Control the aspiration procedure and the desired vacuum via the additional air opening
on the fingertip (fig. 12).
3.
Decommission the device after the aspiration procedure according to section 3.4.
In any case, please take note of section 2.4 and the important information on the
handling of the bacterial filter! Ensure that the battery has a sufficient capacity
during the aspiration procedure. The battery status display must signalize green or
orange. If the display turns red, the power supply unit must be plugged in
immediately. Only in this case the aspiration can be continued without any
interruption.