background image

USER MANUAL

               

Community & Homecare Bed Range

Содержание Community & Homecare Bed Range

Страница 1: ...USER MANUAL Community Homecare Bed Range...

Страница 2: ...g this bed from our Aspire Community and Homecare bed range This product has been specifically designed to offer comfort support and safety It is important that you read this user manual carefully bef...

Страница 3: ...erview 4 Unpacking Inspection 5 Assembly Instructions 6 Operation 7 Accessories 8 Transport Storage 9 Troubleshooting 10 Cleaning 11 Service Maintenance 12 Specifications 13 Technical Information 14 W...

Страница 4: ...here concerns exist around patient safe use Key Features Maximum safe working load SWL of the bed is 220kg with a maximum user weight of 180kg Electrically adjusted height range of 220 710mm Total loc...

Страница 5: ...procedures or conditions which if not followed correctly could result in equipment damage or failure GENERAL Electrical equipment can be hazardous if used incorrectly Residual current devices RCD are...

Страница 6: ...tore the handset when not in use to avoid accidental operation of the beds functions The handset can be stored by hooking it on to the side rail of the bed ALWAYS ensure that all drips drains and powe...

Страница 7: ...identification SELF HELP POLE Handle mount with locator mechanism HEAD SECTION Raises the patient s upper torso to a sitting position FOOT END Marked for ease of identification MATTRESS PLATFORM mattr...

Страница 8: ...ccurs on 1300 133 120 Accessories supplied separately Box 1 of 3 A Head End with lift actuator B Foot End with lift actuator both clearly marked with stickers as HEAD and FOOT Box 2 of 3 C Mattress Pl...

Страница 9: ...sembly Step 1 Lay the two mattress platforms on a flat surface Step 2a Insert connector rods into mattress sections Step 2b Position the frame on its side and insert fastening hand wheels Step 3a Alig...

Страница 10: ...the head end of the bed Step 4a Align foot end identified by sticker with the foot end of the mattress platform Step 4c Tighten hand wheels Step 4b Loosen the fastening hand wheels and insert exposed...

Страница 11: ...een FOOT sticker on the Actuator and Bed Frame and fasten with locking pins Step 7a Locate the FOOT Actuator Step 7d Match the red BACK sticker on the Actuator and Bed Frame and fasten with locking pi...

Страница 12: ...le into sleeve at head of bed aligning notch with groove Step 10a Step 10b Locate side rail sliding mechanism which includes three connector caps Mount connector caps over side rail ends and push unti...

Страница 13: ...ionality for patients with safe use concerns when unsupervised Ensure the key is inserted to the on position prior to use Power on light indicates system is active Backrest incline decline Backrest kn...

Страница 14: ...essed the patient regarding their suitability for Trendelenburg positioning Hand Control Functions continued Bed raise to 1100mm Bed lower to 710mm Measurements taken from top of bed frame to ground l...

Страница 15: ...on the lever at each castor To unlock the brake use your foot to push up on the lever at each castor Self Help Pole This should be placed on the side of the bed that the user gets in out of bed to hel...

Страница 16: ...ng upwards release by pulling green spring button at either end of rail Refer to page 12 for further instructions IV Pole BEA016592 Inserted into the mount brackets at either side of the mattress plat...

Страница 17: ...ter removing one hand wheel from each side of the bed It is important that one fastening hand wheel remains in place on each side so that connector rods remain in place 3 Insert Head and Foot ends int...

Страница 18: ...ITIONS Aspire Community Homecare Beds should be stored in a clean dry well ventilated area The following limits apply during transport and or a storage period of up to 15 weeks duration Ambient Temper...

Страница 19: ...e checked for correct insertion and or damage Lift Actuator control not working TiMotion Electronics Fault All power leads and plugs should be checked for correct insertion and not visible damage Conn...

Страница 20: ...detergent or the facility s recommended cleaning solution It is recommended that the cloth be a white material to prevent bleeding of dye colours to the bed Take extra care in areas that can harbour d...

Страница 21: ...help pole Is the hand control functioning as it should Are the castors brakes and accessories functioning as they should 1 Overall functional check that all hand control functions are operating the be...

Страница 22: ...r braking castors rotate the brake is not fully effective 2 Check the castors for correct operation 3 If any are not working correctly contact an approved service agent Side Rails if fitted 1 The lock...

Страница 23: ...er Body 718mm Highest 1100mm Lowest 710mm 156mm B A C Leg 510mm Main Actuator 706mm 2020 2120mm 990mm Thigh 349mm Sacral 217mm Upper Body 718mm Highest 1100mm Lowest 710mm 156mm B A C D S W L 150 kg T...

Страница 24: ...utput 24V DC max 70V DC Operating Frequency 2 min on 18 min off Push force Pull force Max 0 N to 4000N Protection IP 54 Insulation III User is not isolated from the ground or frame structure Design Fo...

Страница 25: ...ts pertaining to the return of any items under warranty Aidacare acknowledges and conforms to all ACCC guidelines around consumer rights For further information contact Aidacare on 1300 133 120 or ema...

Страница 26: ...2060 Frame Leg Calf Section GSP392065 Mattress Platform Back Section GSP392070 Mattress Platform Leg Section GSP392075 Mattress Platform Calf Section GSP392080 Mounting Hardware Mattress Platform Back...

Страница 27: ...AIDACARE PTY LTD 4 Noonan Road Ingleburn NSW 2565 1300 133 120 www aidacare com au MADE IN VIETNAM ASUM09 3 021116...