TABLE OF CONTENTS:

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1

Package Contents  . . . . . . . . . . . . . . . . . . . . . . . .1

Safety Precautions  . . . . . . . . . . . . . . . . . . . . . . . .2

Setting Up the Unit  . . . . . . . . . . . . . . . . . . . . . .3-9

Figure 21

Front & Side View

 . . . . . . . . . . . . .7

Figure 22

Back & Side View

 . . . . . . . . . . . . . .8

Figure 23

Back View

 . . . . . . . . . . . . . . . . . . . .9

Operation Functions  . . . . . . . . . . . . . . . . . . .10-16

Customizing Electric Motor Settings  . . . . . . .12-13

Operation  . . . . . . . . . . . . . . . . . . . . . . . . . . .17-18

Sterilization & Maintenance . . . . . . . . . . . . . .19-21

Repacking Instructions . . . . . . . . . . . . . . . . . . . .22

Troubleshooting  . . . . . . . . . . . . . . . . . . . . . . . . .23

Replacement Parts List  . . . . . . . . . . . . . . . . . . .24

Symbol Definitions  . . . . . . . . . . . . . . . . . . . . . . .25

Specifications  . . . . . . . . . . . . . . . . . . . . . . . . . . .26

Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27

Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .28-29

P.O. Box 1548 • Woodinville, WA 98072

8333 216

th

Street S.E. • Woodinville, WA 98072

(425) 487-3157 • (800) 426-5913

www.aseptico.com • [email protected]

i.

This device has been tested and found to comply with the

emissions requirements of IEC 60601-1-2:2001-09. These

requirements provide reasonable protection against harmful

electromagnetic interference in a typical medical installation.

However, high levels of radio-frequency (RF) emissions from

electrical devices, such as cellular phones, may disrupt the

performance of this device. To mitigate disruptive

electromagnetic interference, position this device away from

RF transmitters and other sources of electromagnetic energy. 

To prevent injury to people and damage to property,

please heed relevant warnings and remarks. They

are marked as follows:

WARNING:

Serious injury or death may result if

ignored.

CAUTION:

Damage to property or the

environment may result if ignored.

NOTE:

Important additional information

and hints.

INDICATIONS FOR USE:

The AEU-525 IFU (TRANSPORT III PORTABLE DUAL VOLT

UNIT 110/220V) / AEU-525S (TRANSPORT III W/SCALER

DUAL VOLT UNIT 110/220V) is a portable self-contained

dental system that is used for endodontic and general

dentistry applications. 

RX: FEDERAL LAW RESTRICTS THIS DEVICE TO

SALE BY OR ON THE ORDER OF A DENTIST

41EJ

MEDICAL -- GENERAL MEDICAL EQUIPMENT

AS TO ELECTRICAL SHOCK, FIRE AND MECHANICAL

HAZARDS ONLY, IN ACCORDANCE WITH:

UL 60601-1: First Edition

- Medical Electrical

Equipment, Part 1: General Requirements for Safety.

ANSI/AAMI ES60601-1:2005/(R)2012 and A1:2012,

C1:2009/(R)2012 and A2:2010/(R)2012

- Medical

electrical equipment—Part 1: General requirements

for basic safety and essential performance.

CAN/CSA C22.2 NO. 60601-1-08 (2013)

- Medical

Electrical Equipment - Part 1: General Requirements

for Basic Safety and Essential Performance (Adopted

IEC 60601-1:2005, third edition, 2005-12), Includes

Corrigendum 1:2011.

Classifications in accordance with

UL-60601-1:

- Class I
- Types B and BF Applied Parts
- Ordinary Protection

- Not suitable for use in the presence of a flammable

anesthetic mixture with air, oxygen, or nitrous oxide.

Note:

For reliable grounding, connect to receptacle

marked “Hospital Grade”.

Note:

The emissions characteristics of this equipment

make it suitable for use in industrial areas and hospitals

(CISPR 11class A). If it is used in a residential

environment (for which CISPR 11 class B is normally

required) this equipment might not offer adequate

protection to radio-frequency communication services.

The user might need to take mitigation measures,such

as relocating or re-orienting the equipment.

Содержание AEU-525 Transport III

Страница 1: ...OPERATION MAINTENANCE MANUAL AEU 525 AEU 525S Portable Dental System Transport III...

Страница 2: ...R USE The AEU 525 IFU TRANSPORT III PORTABLE DUAL VOLT UNIT 110 220V AEU 525S TRANSPORT III W SCALER DUAL VOLT UNIT 110 220V is a portable self contained dental system that is used for endodontic and...

Страница 3: ...rate manuals may be provided to cover the operation and maintenance of other accessories for your unit PACKAGE CONTENTS Portable Case with Electronic Control Compressor and Vacuum Pump AE 230M 40 Moto...

Страница 4: ...event the proper function of cardiac pacemakers Therefore Aseptico recommends that patients with a cardiac pacemaker should not be treated with the Ultrasonic Scaler component of the AEU 525S System W...

Страница 5: ...the outside of the rear metal door Fig 4b Turn fastener counter clockwise 1 4 turn to loosen Stand unit upright Use the fastener tool to loosen the two fasteners on the small metal door located on th...

Страница 6: ...9 with Step 8 for easier setup of the syringe and optional scaler Place air water syringe and optional scaler into their respective instrument holders on the air electric module refer to the location...

Страница 7: ...to place by hand tightening the hold down thumb screw on the housing until snug Fig 13b 15 Ensure that the lid on the waste container is securely latched onto the container Hang the container onto the...

Страница 8: ...s as a circuit breaker for the whole unit The right hand switch acts as a circuit breaker for the compressor Electrical overloads will trip the two switches If either switch is found in a neutral posi...

Страница 9: ...CONTROL WASTE CONTAINER LCD DISPLAY SALIVA EJECTOR SYRINGE SCALER ELECTRIC MOTOR ELECTRIC MOTOR CONTROL PANEL Pages 10 13 MANIFOLD PANEL Pages 14 15 SCALER CONTROLS HANDPIECE WATER CONTROLS PRESSURE...

Страница 10: ...OTTLE HVE HOSE SETTING UP THE UNIT Cont d AIR ELECTRIC MODULE ASSEMBLY FOOT CONTROL HIGH VACUUM HOSE WATER BOTTLE PURGE SWITCH SALIVA EJECTOR HOSE LOW VACUUM HOSE INSTRUMENT HOLDERS Page 10 ON OFF LEV...

Страница 11: ...ULL ALARM LED Page 16 WATER BOTTLE PURGE SWITCH Page 6 COMPRESSOR ON OFF SWITCH CIRCUIT BREAKER POWER ENTRY AND CIRCUIT BREAKER PANEL Page 6 115V 230V VOLTAGE SELECTOR SWITCH POWER INLET ON OFF SWITCH...

Страница 12: ...ower On Off to the Transport III When the unit is turned On the green LED next to the power switch will illuminate 2 COMPRESSOR CIRCUIT BREAKER SWITCH Fig 24 When the switch is On circuit closed conti...

Страница 13: ...to turn on the control panel and return the motor to its last used settings 2 Press the RATIO button repeatedly to select the ratio that matches the handpiece being used Available ratios are 1 5 1 2 1...

Страница 14: ...ILLUMINATION button to turn the hand piece LED On Off To activate the handpiece and LED remove the handpiece from its holder turn the Illumination button On and press the foot pedal to operate the mo...

Страница 15: ...FWD 100 Off 2 60 000 FWD 100 Off 3 60 000 FWD 50 Torque Limit Off 4 40 000 FWD 100 Off 5 40 000 FWD 50 Torque Limit Off 5 1 1 2 000 FWD 100 On 2 1 500 FWD 100 On 3 1 200 FWD 100 On 4 800 FWD 100 On 5...

Страница 16: ...bgingival scaling furcations supragingival fine scaling and spot removal The 38 tip is used for lingual and buccal subgingival scaling and furcations The 39 tip is used for universal lingual and bucca...

Страница 17: ...rd position to turn handpiece water On or downward to turn it Off A water control valve is also provided to adjust the volume of water flow to the handpiece This valve is located next to the toggle va...

Страница 18: ...Low Volume Saliva Ejector compressor system from running excessively ensure that the vacuum heads are firmly cradled in their holders after each use and that their control levers are toggled Off beep...

Страница 19: ...er the Transport III unit has been set up and you have made yourself familiar with the operation functions you are ready for operation as follows 2 Turn the power and circuit breaker switches on the r...

Страница 20: ...b Adjust the water and air flow by using a screwdriver to turn the two small adjustment screws located on the side of the air electric module Upper screw controls water lower screw controls air Turn...

Страница 21: ...assembly at 132 C 270 F for ten minutes Loosely coil the motor cord when autoclaving Avoid sharply bending the cord when autoclaving Alternatively wipe down the motor cord with disinfecting solution...

Страница 22: ...aming cleansers in the vaccum lines AIR BOTTLE FILTER Routinely check the air bottle filter once a day for condensation To drain condensation place a towel or container below the filter and use pliers...

Страница 23: ...ips are fully autoclavable Disinfect and clean the cover and tips before autoclaving Autoclave at a maximum temperature of 135 C 275 F for 10 minutes or 120 C 248 F for 20 minutes Wipe off the scaler...

Страница 24: ...wn in Fig 46 10 Coil the two power cords and position on top of the waste tank and air bottle as shown in Fig 46 11 Carefully close case lid taking care not to pinch any internal components 12 Locate...

Страница 25: ...r flow control valve to the handpiece is open counterclockwise Check that handpiece air toggle on side of air electric module is On Check that the air flow control valve to the handpiece is open count...

Страница 26: ...100 MICRON WATER BOTTLE 730326 1 BOTTLE 1 LITER 32 OZ BLOW MOLDED HDPE WATER 730471 1 GASKET FOR AIR WATER BOTTLE LID 730473 3 FILTER AIR 1 4 NPT X 1 19 GAST INLET FILTER 730494 1 BOTTLE 750 ML 25 OZ...

Страница 27: ...tion Atmospheric Pressure Limitation Type BF Equipment Motor Direction Light Controls Consult Instructions For Use Do Not Lift by Top Lid Or Latches Counterclockwise to Increase Open Clockwise to Decr...

Страница 28: ...33 9 fl oz 1 0 liters Air Storage Capacity 25 4 fl oz 750 ml nominal Water Flow 5 07 fl oz min 0 15 liter min Waste Container Capacity 0 48 gallon total 0 24 gal per side 1 8 liters total 0 90 L per...

Страница 29: ...ter of such action In the event of alleged defect under warranty the purchaser is to notify Aseptico s Customer Service Department promptly Customer Service will provide instructions usually directing...

Страница 30: ...28 NOTES...

Страница 31: ...29 NOTES...

Страница 32: ...P N 420996 Rev K ECO 15011 06 2020 PRINTED IN THE U S A P O Box 1548 Woodinville WA 98072 8333 216th Street SE Woodinville WA 98072 425 487 3157 800 426 5913 www aseptico com info aseptico com...