DrillSaw Sports 400™ System
User’s Guide
DFU-0253-4r0_fmt_en-US
Page 16 of
46
3.4
Electromagnetic Compatibility (EMC)
This device has passed testing for EMI/RFI radiation and susceptibility
and EMC compatibility. This device may cause interference with other
devices in the near vicinity if not set up and used as instructed by
Arthrex.
The DrillSaw Sports 400 System has been designed to accept EMC from other devices
within the limitations as described in section 11.0.
Try to correct the interference by following one or more of these measures:
1.
Reorient or relocate the receiving device.
2.
Increase the distance between the devices.
3.
Connect the device to an outlet on a different circuit than the other
device(s) are connected to.
4.
Consult the manufacturer or field service technician for the receiving
device for guidance.
3.5
Risks due to electromagnetic fields:
The functionality of implantable systems, such as cardiac pacemakers and ICD
(implantable cardioverter defibrillator) can be affected by electric, magnetic and
electromagnetic fields.
1.
Consider the application and find out if patient or user has any of these
devices, before using the product.
2.
Weigh the risks and benefits.
3.
Keep the product away from any of these devices.
4.
Make appropriate emergency provisions and take immediate action upon
any signs of health issues.
5.
Symptoms such as raised heartbeat, irregular pulse and dizziness can be
signs of a problem with a cardiac pacemaker or ICD.