5.5 Parameter adjustment
5.5.1 Changing patient airway pressure during therapy
In the CPAP and helmet CPAP modes, the patient airway pressure can be
adjusted by changing the PEEP valve during the treatment.
In the CPAP and helmet CPAP modes, the patient airway pressure field
shows the measured patient airway pressure, which is averaged over a
longer period of time. To view the instantaneous change of the patient
airway pressure touch the patient airway pressure field.
Note
In the CPAP and helmet CPAP modes, the patient airway
pressure change can also be displayed without touching the
patient airway pressure field. However, the update of the
average value will take considerably longer.
Содержание AquaVENT FD140
Страница 1: ...AquaVENT FD140 Instructions for use CRITICAL CARE...
Страница 6: ...0 Contents 4 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 11: ...Instructions for Use AquaVENT FD140 Rev 12 9 2Safety In this chapter 2 1 Safety precautions 10...
Страница 16: ...2 Safety 14 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 56: ...5 Operation 54 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 66: ...6 Alarms and messages 64 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 70: ...7 Cleaning 68 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 76: ...8 Maintenance 74 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 88: ...9 Technical specifications 86 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 95: ...Instructions for Use AquaVENT FD140 Rev 12 93 This page is intentionally blank...