4 Preparing the device for operation
28
Instructions for Use, AquaVENT
®
FD140, Rev. 12
4.6 CPAP or Paed CPAP with face mask, heated
An active humidifier must be integrated in the heated CPAP breathing
circuit.
Note
To prevent water from the humidifier, which collects during
operation, from entering the AquaVENT
®
FD140, the active
humidifier should be placed at a level below that of the
AquaVENT
®
FD140.
Note
When removing the active humidifier, first disconnect the tube
from the AquaVENT
®
FD140 to prevent water from the
humidifier from entering the AquaVENT
®
FD140.
Содержание AquaVENT FD140
Страница 1: ...AquaVENT FD140 Instructions for use CRITICAL CARE...
Страница 6: ...0 Contents 4 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 11: ...Instructions for Use AquaVENT FD140 Rev 12 9 2Safety In this chapter 2 1 Safety precautions 10...
Страница 16: ...2 Safety 14 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 56: ...5 Operation 54 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 66: ...6 Alarms and messages 64 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 70: ...7 Cleaning 68 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 76: ...8 Maintenance 74 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 88: ...9 Technical specifications 86 Instructions for Use AquaVENT FD140 Rev 12 This page is intentionally blank...
Страница 95: ...Instructions for Use AquaVENT FD140 Rev 12 93 This page is intentionally blank...