5.2 Error messages
AE-4022
5-3
5.2
Error messages
Errors may occur when you operate the instrument incorrectly or make a mistake during normal operation.
When an error occurs, an error message appears on the screen and an alarm (pi-pi-pi-pi) sounds for approximately 1
minute.
Press
to cancel the error notice. The display returns to the [Standby screen].
To prevent the error from recurring, take appropriate measures, referring to the table below. Alternatively, note the
details of the situation at hand, turn OFF the instrument, and contact your local distributor.
Message
Error
Possible cause
Measures
E001
Power down
The instrument was suddenly turned
OFF during measurement.
Retest the sample(s) currently being
measured.
E002
Backup memory error
As the instrument was unused for a
long period, the memory backup
battery was exhausted and the
measurement results stored in the
memory were lost.
Keep the instrument powered for at
least 11 hours to charge the backup
battery. After charging the instrument,
set the date and time. (See page 3-11.)
E004
No paper in the printer
• The thermal recording paper has
run out.
• The thermal recording paper roll
was not properly installed.
• Install a new roll of thermal recording
paper. (See page 4-12.)
• Make sure the thermal recording
paper is installed correctly.
E005
Waste box is full
At the end of the measurement, the
test strip waste counter exceeded 90.
Cancel the error notice by pressing
. Empty the used test strips from the
waste box, and press
.
E006
Surplus urine is full
At the end of the measurement, the
total number of measurements since
turning ON the instrument exceeded
190.
Turn OFF the instrument and discard
any surplus urine.
E007
Data not found
No data was found in the specified
range.
Verify that the specified range for
re-printing, re-sending, or list printing is
appropriate.
E008
Auto start sensor error
• A test strip was placed while the
carrying arm was moving to the
suction ports.
• The detection section of the auto
start sensor was soiled with urine.
• The auto start sensor is faulty.
• Remove the test strip.
• Clean the detection section.
• If you find no abnormality and the
error recurs, contact your local
distributor.
E009
No test strip on the
feeder
• The feed lever is not attached.
• The incoming strip sensor is faulty.
• The sensor failed to detect a test
strip.
• Turn OFF the instrument and open
the maintenance cover.
• Mount the feed lever.
• If you find any obstacles or scattered
test strips inside, remove them.
• If you find no abnormality and the
error recurs, contact your local
distributor.
E011
Elapse of QC deadline
The QC deadline elapsed while the
instrument was OFF or on standby.
Perform control measurements.
Содержание AUTION ELEVEN AE-4022
Страница 1: ......
Страница 75: ...Chapter 2 Measurement Operation AE 4022 2 40 ...
Страница 112: ...3 7 SETUP user settings AE 4022 3 37 4 Quit the SETUP menu aPress twice The display returns to the Standby screen ...
Страница 143: ...Chapter 4 Maintenance AE 4022 4 14 5 Close the printer cover aClose the printer cover Printer cover ...
Страница 146: ...AE 4022 5 1 Chapter 5 Troubleshooting 5 1 Warning messages 5 2 5 2 Error messages 5 3 5 3 Trouble messages 5 5 ...
Страница 153: ...Chapter 5 Troubleshooting AE 4022 5 8 ...
Страница 160: ......
Страница 161: ......