MAGS1 User Guide
25
V1.5 22-1-2016
10. MAGS1 Error Code Table
MAGS1 can detect and show a number of errors in one error code value.
Error position
Error Description
0
Empty Pipe (Air Detect)
1
Overloaded
2
Excitation
3
– 6
RESERVED (non-use)
7
AD converter error
8
– 12
RESERVED (non-use)
13
Temperature error
14-15
RESERVED (non-use)
Errors in Modbus register are indicated in hex format and have to be converted to binary
format!
Once the error code has been converted to binary format, each position is related to a different error
(see the table above). Number 1 indicates error and number 0 indicates no error.
Example:
Error shown in Modbus
register:
Error position:
Read errors:
76543210
085HEX =
10000101 BIN
AD converter / Excitation / Empty pipe
Position 0
Position 1
Position 15