Verification of the Charger’s Power
Source
If the light does not illuminate when the battery is
correctly installed in the
Voyager
charger, try the
following:
1) Make sure that the charger is correctly
plugged into the AC outlet, and that the green
light is on.
2) Make sure that the battery is properly inserted
in the charger.
3) Check the power of the AC outlet on the wall.
4) If the charger’s yellow or green light does not
light up, contact your local Joerns / Healthcare
agent for assistance.
Sling Inspection and Care
Refer to documentation included with slings.
Annual Inspection
The
Voyager
and its accessories must be
inspected annually by a qualified technician.
UK Regulation
WARNING: The slings should be checked
before and after use and, if necessary,
washed as per instructions on the sling.
This is especially important when using the
same equipment for another patient. This
minimizes the risk of cross infection.
WARNING: The
Voyager
and accessories
must be serviced every 12 months as a
minimum requirement. Do not attempt to do
the inspection unless you are qualified to do
so. Not servicing the product may lead to
patient fall, causing injuries.
Note: In the UK the
Voyager
must be
inspected and maintained in compliance
with 1998 no 2307 health & safety: lifting
operations and lifting equipment regulations
1998. Your dealer will provide you with more
information.