66
Instructions For Use
EN
Save Local Settings = Fix all
current settings as defaults.
Apply Defaults = Overwrite all
current settings with the stored
defaults.
Imports and Exports allow transfer
to and from Team3 via USB Flash
Drive.*
Note: Do NOT transfer settings files from one monitor to another if the software versions are not the same.
Note: Import/Export buttons are only displayed when a USB Flash Drive is installed.
*A software tool is available from your supplier to support the Easinote function. See section 13.6.
Allows NiBP protocol details to be
customised.
13.6 Customising Easinotes
WARNING
The Secure Settings menus should be accessed by authorised personnel only.
The Easinote sub-system is very flexible allowing up to 12 groups of 12 notes to be configured on Team3.
Refer to Section 13.5 for details on how to change the Language and to access the Settings Management
feature.
Exporting Easinotes
1. With Team3 switched on, plug a USB flash drive into one of the Team3 USB sockets.
2. Set up Team3 to operate in the language of your choice. This will ensure that the Easinote file to be
exported is saved with the other files for that language. It is important to remember that each language has
its own Easinotes file so Easinotes will change with Language selection.