13. Maintenance
Warning
It is very important that all instructions in the Maintenance section are followed carefully.
13.1 User maintenance
The checks below can be performed by any user of the equipment.
Mechanical inspection
Every three months:
1. Inspect the AC supply cable, transducers, and all other assemblies and connectors for loose or broken
parts, or any other damage.
2. Pay particular attention to the AC supply socket.
3. Look carefully for cracks which may allow the ingress of liquids or gels.
4. Replace any broken or damaged transducers or cables.
5. If there is damage to the main Team3 unit, contact your local Huntleigh Healthcare Ltd representative.
Cleaning the print head on the printer
1. Pull the paper tray out as far as it will go.
2. Remove the paper pack.
3. Using a lint-free cloth and pure alcohol, wipe along the full width of the print head, which is beneath the
plastic edge of the paper compartment.
4. Replace the paper tray and paper pack.
Check NIBP cuffs and hose
Once a month:
1. Check the NIBP hose. Straighten out any kinks and distortions.
2. Check the cuff(s) for wear and damage.
13.2 Technical maintenance
Refer to your Service department for details of technical maintenance and support.
13.3 Corrective maintenance
All corrective maintenance must be performed by qualified engineers approved by Huntleigh Healthcare Ltd.
The Sonicaid Team3 Service Manual (order part number 777490) is designed as an aid to engineers in
maintenance and service of repairable parts.
13.4 Servicing
Servicing should be performed only by Huntleigh Healthcare Ltd or their appointed service agent. If you have
difficulty obtaining service for Team3, contact Huntleigh Healthcare Ltd.
WARNING
Servicing cannot be performed while the unit is in use.