QUICK REFERENCE GUIDE – CLEANING AND DISINFECTION

Caregiver obligations

•  To avoid eye and skin damage, always use protective glasses and gloves. If contact occurs rinse with plenty of water. 

If eyes or skin becomes irritated, contact a physician.

•  To prevent cross-contamination, always follow the disinfection instructions in the Instructions for Use.

•  To avoid eye or skin irritation, never disinfect in the presence of a patient.

•  Always read the instructions of the disinfectant. 

Clean and disinfect with spray handle or spray bottle

  

(Still bath and Hydrosound system only.)

Remove visual residue

1.  Check the level of disinfectant in the disinfectant bottle.

2.  Make sure all functions are turned off.

3.  Open the drain to empty any left over water.

4.  Detach and place accessories at the bottom of the bathtub.

5.  Press the Patient Shower button on the control panel.

6.  Shower and rinse the bathtub and the accessories to remove visible 

residue. Turn the accessories and rinse again.

Clean

7.  Put on protective gloves and protective glasses.

8.  Open the access door on the left side of the panel and lift out the 

disinfection spray handle or use a disinfection spray bottle.

9.  For disinfection spray handle only:

 Direct the disinfectant spray towards 

the bottom of the bathtub and press the Disinfection button on the control 
panel. 

10. For disinfection spray handle only

: Adjust water temperature so it 

reaches approximately 25 °C (77 °F) to eliminate disinfectant fumes.

11. 

6SUD\GLVLQIHFWDQWLQWKHEDWKWXEDQGFKHFNGLVLQIHFWDQWPL[UDWLRRQÀRZ

meter to make sure it is within the stated range (see Instructions for Use).

12. Spray disinfectant in the drain.

13. Scrub the drain with a long handled brush.

14. Close the drain.

15. Spray disinfectant on all internal bathtub surfaces and accessories. Pay 

attention to knobs and sensors attached to the bathtub surface such as 

GUDLQFRQWURONQREDQGVXUIDFHRYHUÀRZRXWOHW

16. 

6SUD\GLVLQIHFWDQWLQWRWKHVXUIDFHRYHUÀRZRXWOHW

17. For disinfection spray handle only

: Press the Disinfectant button on the 

control panel to turn it off, and place the disinfectant spray handle back 
behind the access door. 

18. Scrub the surface of the bathtub, loose parts and accessories with a soft 

bristle brush or cloth. Scrub thoroughly, especially on the handles and hand 
controls.

19. Open the drain.

20. Press the Patient Shower button on the control panel.  

21. Rinse all parts clean with plenty of water having a temperature of 

approximately 25 °C (77 °F) to remove disinfectant.

22. Soak cloth in a disinfectant. 

23. Wipe all other contact areas with the cloth, such as the outside of the bath, 

sensors and hand control. 

24. Wet a new cloth with water and remove all traces of disinfectant on contact 

areas. Rinse the cloth occasionally with running water. 

25. Examine the bathtub and accessories to make sure that all of the 

disinfectant is removed. Repeat rinse if necessary.

26. Examine bathtub and accessories for visible damage. 

Disinfect

27. Open the access door on the left side of the panel and lift out the 

disinfection spray handle or use a disinfection spray bottle.

28. For disinfection spray handle only

: Direct the disinfectant spray towards 

the bottom of the bathtub and press the Disinfection button on the control 
panel. 

29. Close the drain.

30. For disinfection spray handle only:

 Adjust water temperature so it 

reaches approximately 25 °C (77 °F) to eliminate disinfectant fumes.

31. Spray disinfectant on all internal bathtub surfaces and accessories. Turn the 

accessories and spray and on the other side. Pay attention to knobs and 
sensors attached to the bathtub surface. 

After using the disinfectant for a period of time, the system stops and is locked 

IRUDSUHGH¿QHGWLPHWRHQVXUHGLVLQIHFWDQWFRQWDFWWLPH7KH/('RQWKH

Disinfection button is lit until the system is unlocked.

32. For disinfection spray handle only:

 Press the Disinfection button on the 

control panel to turn it off, and place the disinfectant spray handle back 
behind the access door. 

33. Allow appropriate contact time according to the instructions of the 

disinfectant.

34. Open the drain. 

35. Press the Patient Shower button on the control panel. 

36. Rinse all parts clean with plenty of water having a temperature of 

approximately 25 °C (77 °F) to remove disinfectant.

37. Soak a cloth in disinfectant.

38. Wipe all other contact areas such as the outside of the bathtub, sensors and 

hand control with the cloth.

39. Allow appropriate contact time according to the instructions of the 

disinfectant.

40. Wet a new cloth with water and remove all traces of disinfectant on contact 

areas. Rinse the cloth occasionally in running water.

41. 

/HDYHWRDLUGU\

42. Attach any accessories. 

Clean and disinfect Century before every use.