Replacing the Battery Pack
•
Press the
I/O
button to turn the power OFF.
•
Unplug the power cord.
•
Use a Phillips screwdriver to remove the four (4) screws securing the battery panel as shown
above.
•
After removing the screws, carefully set the panel to the side. Note that it is still connected to
the instrument via the power connector.
•
Slide the battery pack out of the J605.
•
Disconnect the battery connector to detach the old battery.
•
Install the new battery by connecting it to the battery connector and sliding it into the J605.
•
Tuck the cabling into the gap between the battery and the case.
•
Replace the battery panel and secure with the four (4) screws removed above.
•
Dispose of the old nickel metal hydride (NiMH) battery in accordance with state and federal
regulations.
•
After battery replacement, the date and time will need to be updated. See the
for directions. Sample data is stored in non-volatile memory and will not be
lost or affected when the battery is disconnected or replaced.