![Arizona Instrument JEROME J605 Скачать руководство пользователя страница 34](http://html1.mh-extra.com/html/arizona-instrument/jerome-j605/jerome-j605_user-manual_2972212034.webp)
AZI Customer Service 800-528-7411 or 602-470-1414
Page 34 of 57
Internal Filters
Replace the internal filters after six (6) months of use, or as needed.
•
Press the
I/O
button to turn the power OFF and unplug the power cord.
•
Remove the two (2) side screws from the intake end of the instrument and open the case.
•
Carefully disconnect the Tygon
®
tubing from both ends of the filters and discard the old
filters.
CAUTION:
Scrubber filters contain Resisorb
®
. Used filters will also contain trace
amounts of hydrogen sulfide. Use proper methods when disposing of used
filters. Call AZI Customer Service at 800-528-7411, 602-470-1414, or e-mail
[email protected] for a copy of the Resisorb
®
MSDS or for other questions.
•
Replace the old filters with new filters, oriented as shown in the diagram above.
Push the Tygon
®
tubing as far as it will go onto the filter fittings.
•
Push the filters into the mounting clips.
•
Remove any crimps or twists in the tubing and ensure that tubing connections are secure. If
the tubing is loose, readings may not be accurate. Replace any tubing that has deteriorated
due to heat and/or age.
•
Close the case and replace the screws.
•
Dispose of all filters in accordance with state and federal environmental regulations.
600-0281 LFD
Scrubber Filter
600-0279 LFS
Scrubber Filter
Z2600 3930
Scrubber Filter
2600 3061
Acrodisc
®
Filter