INSTRUCTIONS

FUNCTION CHECK CARUM BREASTPUMP 

The Carum breastpump is maintenance-free. Ardo medical AG recommends that operational and vacuum functions 
be checked by trained personnel at least every two years.

Vacuum Control

1

Required Test Equipment 

Ardo Manometer (Art. No. 63.00.113)

Ardo Pumpset breast shell size 
26 mm (e.g. Clinic Pumpset Art. No. 
63.00.42).

2

Procedure

Connect the Pumpset to the Carum 
unit and insert the plug of the mano-
meter fi rmly and securely into the 
breast shell.

Switch on the unit by pressing the  
On/Off key.

The unit will be in stimulation mode 
(orange background).

00:20

Vacuum

Cycles

CARUM

00:00

Vacuum

Cycles

CARUM

Press the "Mode" key to switch from 
stimulation mode to express mode 
(green background) .

00:00

Vacuum

Cycles

CARUM

02:00

Vacuum

Cycles

CARUM

In express mode, increase the vacuum 
to maximum. Read the vacuum level 
value on the manometer.

02:26

Vacuum

Cycles

CARUM