Besides this an audible alarm activates. After mains power is connected the pump can be
restarted. Alarms are cancelled automatically as soon as mains power is restored and the
green pictogram in the information window is illuminated:
A low battery alert is activated approximately 30 minutes before the battery is depleted. The
alarm symbol flashes and an audible alarm is activated.
To silence the audible alarm, press the “Alarm Silence” key (SELECT key). The battery
symbol continues to flash until the pump is reconnected to the mains.
Warning: It is not allowed to use a pump with a defect battery or a battery which does
not pass the battery test (point 8 of safety check
– chapter 9.1) on a patient. In case of
a mains failure the pump might switch off and stop infusing without any alarm or
warning. Defect batteries or batteries which do not pass the battery test (point 8 of
safety check
– chapter 9.1) must be replaced..
Warning: For safety reasons only batteries supplied by
arcomed ag
may be used as
replacement.
4.5
Air in Line Alarm
The air in line detector can detect air bubbles of a size of 50
l and bigger. Not every air
bubble causes an air in line alarm. If the pump is operated with default settings, air bubbles
smaller than 150
l will not cause an air in line alarm.
The volume of all detected air bubbles is added up continuously. If more than 1000
l of air
pass the air in line detector within 15 minutes a cumulated air alarm is triggered. This way it
can be assured that a large number of small air bubbles (smaller than 150
l) will not harm a
patient.
If an air alarm is triggered the pump stops and an acoustic alarm is emitted. The red alarm
pictogram is illuminated.
To rectify the air alarm it is necessary to disconnect the patient and prime the set again.
4.6
Occlusion Alarm
The Volumed® µVP7000 Chroma has automatic pressure monitoring whereby the pressure
in the system is calculated by measuring the hardness of the tubing with the pressure
transducer. If the pressure rises above the set occlusion pressure limit the pumps stops and
emits an acoustical and visual alarm.