Operating Instructions

OS3 Base unit, VC830100

VV016011E

Phaco Emulsification System

0297

Содержание OS3

Страница 1: ...29 05 09 VV016011E 1 Operating Instructions OS3 Base unit VC830100 VV016011E Phaco Emulsification System 0297...

Страница 2: ...taltic operation 13 8 2 3 Infusion pressure IOP 13 8 2 4 Continuous irrigation 14 8 2 5 Reflux 14 8 2 6 Proportional Override 14 8 3 Ultrasonic phaco 14 8 3 1 Phaco test 14 8 3 2 Phaco operation 14 8...

Страница 3: ...may be impaired or the pacemaker itself damaged If in doubt the cardiology department must be consulted 2 The functioning of other electro medical equipment may be affected during the operation of bip...

Страница 4: ...k green arrow buttons immediately beneath the display The lower half of the control panel houses the displays and adjustment buttons for the irrigation aspiration system The lowest row on the control...

Страница 5: ...g factor can be changed see 8 3 4 PULSE button Switches on pulse or burst operation for phaco 1 2 3 and switches it off when the button is pressed again If the button is held down the pulse or burst p...

Страница 6: ...tween positions 0 1 2 and 3 If the pedal is pressed slowly the settings 0 3 can be felt as a slight resistance This resistance can be changed in ParaProg low medium high The pedal is brought to positi...

Страница 7: ...ear Linear instrument control Aspiraton Phaco VIT DIA LOW linear CAPS REGULAR fix VISCO INJECTION linear Reflux Instant diathermy Pulse on off linear phaco Pos 3 Instrument Phaco VIT DIA CAPS Override...

Страница 8: ...the installed software appears in weakly illuminated lettering The unit is now ready and can be prepared for the operation 5 1 Replacing fuses Press the clip on fuse holder P to the right until it pop...

Страница 9: ...trument connection tubes must be kept in the sterile area The sterile person must not insert the tubes into the unit Turn the cassette rotary switch in an anti clockwise direction and open the cassett...

Страница 10: ...P button twice 6 3 Removing and cleaning the cassette 6 3 1 Cassettes for single use Fig 3 4 Remove the tubes from the instruments and join together connections 1 and 2 to prevent the liquid from drip...

Страница 11: ...g the irrigation tube i and the aspiration tube a into the I A handpiece The luer connectors are designed so that it is impossible to connect the instruments wrongly Attach the chosen tip Wet the tip...

Страница 12: ...CO1 PHACO2 PHACO3 CMP1 CMP2 CMP3 as well as VIT A VIT B and VIT PN or VIT A1 VIT A2 etc Calling up values Press the above buttons briefly to call up the stored values immediately Storing new values Pr...

Страница 13: ...03 or VV 630010 are being used you can change between the venturi and peristaltic pump at any time To do so the pedal must be in the zero position Effect the change by pressing the venturi button on t...

Страница 14: ...be tested before beginning the operation To do so fit a filled test chamber to the handpiece and press the TEST button Never carry out the TEST procedure whilst the handpiece is in the eye NOTE The te...

Страница 15: ...ional to the pedal deflection Note CMP1 2 3 have their own memories By holding down the corresponding button phaco 1 2 3 see 8 1 for a short period all settings will be memorized 8 3 5 Occlusion Mode...

Страница 16: ...ains available without having to set the unit to diathermy mode by pressing the DIA button This function can be set to DiaLOW or DiaHIGH or switched off in ParaProg NOTE ParaProg can be used to determ...

Страница 17: ...nd 3 Duallinear vitrectomy This operating mode enables separate control of the pump by vertical pedal movement and control of the cutting rate by moving the pedal horizontally to the left or vice vers...

Страница 18: ...es Every surgeon develops his own preferred operating technique which also requires corresponding specific setting values for the various stages of the operation The OS3 unit can optimally accommodate...

Страница 19: ...are temperature 134 C 138 C min cycle time 3 minutes When the instruments are removed from the sterilisation system they must be cooled to room temperature before commencing an operation The instrume...

Страница 20: ...itrectomy 200Vss at instrument input Aspiration flow 0 50 ml min 25 Vacuum 0 600 mmHG 10 0 80kPa 10 Noise level 70 dB A Dimensions 380 x 150 x 340 mm W H D Weight Unit 14 8 kg Pedal 4 8 kg Transport a...

Страница 21: ...heel System ready The self test has been completed successfully Check tip The tip is not properly attached See Section 8 3 1 Call Service Error 1 Contact service Internal voltages defective The unit c...

Страница 22: ...aco accessories included Vitrectomy accessories included 18 Calibration and maintenance Provided that the cassettes are replaced regularly in accordance with the instructions in the packaging this uni...

Страница 23: ...nnection NIST EN 397 L Slot for smart card see service manual M1 Pedal port M2 Reserve port M3 Control panel port N Serial interface see service manual O Mains connection socket P Fuse holder Q Mains...