THD:
Total harmonic distortion for the measured
signal.
K Factor:
K-factor, which is a measure of the heating ef-
fect of harmonic energy in magnetic materials.
Harmonic
Amplitude:
Amplitude in Volts or Amps of harmonic
2 through 50.
Harmonic
Phase:
Phase angle of harmonic 2 through 50 with
respect to the fundamental.
3.11.5.6
Waveform Data
The Model 931A will export waveform data while the print operation is
executed. This occurs when performing measurements in the waveform
display mode, or reviewing stored data while in that mode.
Waveform data is tabular format and similar to harmonic data. It
consists of 240 records, which correspond to points on the waveforms for
measurement channels 1 and 2.
Following is the format for waveform data:
log data #, date, time, channel 1, channel 2
C
“”, ””, ””, channel 1, channel 2
C
“”, ””, ””, channel 1, channel 2
C
. . .
C
. . .
C
“”, ””, ””, channel 1, channel 2
C
Log Data #:
A number assigned to the data record.
If
the data was stored prior to print/export, this
number will correspond to the record number
assigned when stored. If the data is printed or
exported directly, this number will always be
zero.
96