Model 125W1000A
24
Rev A
2.10.8
Operating Hours (RF ON) Query
This command queries the total hours spent with the RF enabled.
Query:
OH?
Response Format:
OH=<hours><LF>
<hours>:
The total hours with RF Enabled (0 to 100000), padded with spaces to always
be 6 digits wide.
Example:
To query the RF On operating hours, send the following command:
OH?<LF>
Response:
OH=<space><space><space><space>37<LF>
(The system has spent 37 Hours in an RF On state)
2.10.9
Operating Hours (Power ON) Query
This command queries hours spent in Power ON.
Query:
OHP?
Response Format:
OHP=x<LF>
<hours>:
The total hours with Power On (0 to 100000), padded with spaces to always
be 6 digits wide.
Example:
To query the Power On operating hours, send the following command:
OHP?<LF>
Response:
OHP=<space><space><space>428<LF>
(The system has spent 428 Hours in a Power On state)
2.10.10
Faults Query
This command queries information about a currently active fault.
Query:
FSTA?
Response Format:
FSTA=<space><status><LF>
<status>:
A hexadecimal number (0000 to FFFF), padded with zeros to always be 4
characters wide.
Example:
To query the fault status, send the following command:
FSTA?<LF>
Response:
FSTA=<space>0003<LF>
(Interlock Fault)
Note:
See the Fault Definition section for a detailed description of all faults.
2.10.11
Reset Faults
This command clears a resettable fault. A fault can only be cleared if the underlying issue that created the fault
has been resolved.
Syntax:
RESET
Parameters:
None
Example
To clear a reported fault, send the following command:
RESET<LF>
Note:
None