![AR 10052381 Скачать руководство пользователя страница 44](http://html1.mh-extra.com/html/ar/10052381/10052381_operating-and-service-manual_2962299044.webp)
Model 125W1000A
22
Rev A
2.10.5
Identity Query
This command queries the amplifier for its identity response.
Query:
*IDN?
Response Format:
AR-RF/MICROWAVE-INST,<model>,<firm rev><LF>
<model>:
The model designation of the amplifier.
<firm rev>:
The firmware revision of the control software.
Example:
To query the identity of the amplifier, send the following command:
*IDN?<LF>
Response:
AR-RF/MICROWAVE-INST,125W1000A,1.0<LF>
2.10.6
IO Board Firmware Revision Query
This command queries the firmware revision of the internal I/O Board.
Query:
*IOB?
Response Format:
INTERFACE_BOARD_SW_REV<firm rev><LF>
<firm rev>:
The firmware revision of the control software.
Example:
To query the firmware of the I/O board, send the following command:
*IOB?<LF>
Response:
INTERFACE_BOARD_SW_REV3.20<LF>
2.10.7
Amplifier State Query
This command queries the state of the amplifier.
Query:
STATE?
Response Format:
STATE=<space><char1><char2><char3><char4><LF>
<char1> - <char4>:
ASCII character (0 to 9 or A to F) representing a hexadecimal character
Example:
To query the state of the amplifier, send the following command:
STATE?<LF>
Response:
STATE=<space>8301<LF>
(Remote Mode, Power On, RF OFF, and Manual Mode)
Each hexadecimal character represents a 4-bit binary number, allowing for a total of 16 binary flags that
can be used to indicate various states of the amplifier. The definition of what each bit of information
represents can be found in the
Amplifier State Response Table
below. For help converting the ASCII
characters into binary bits, see the
Hex character to Binary Bits
table.