Model 5000A225A
32
Rev B
2.4.2.6.15
Operating Hours (Power On)
Query to get the Power On operating hours.
Syntax:
OHP?
Parameters:
None
Response Format:
OHP=x<LF>
Where:
x
=
0
to
100000
Units are Hours. Values can be up to six digits in length. Leading zeros are read as
spaces.
Example:
To find out the Power On operating hours, send the following query.
OHP?<LF>
Response:
OHP=<space><space><space>428<LF>
(The system has spent 428 Hours in a Power On state)
2.4.2.6.16
Initiate MPA Coolant Temperature and Transistor Currents
Command to initiate the process of acquiring the MPA coolant temperature and transistor currents.
Syntax:
MPAn
Parameters:
n
Where n is the two digit MPA number which can only be 01.
Response Format: None (No Query for this command)
Example:
To initiate the process of finding out the coolant temperature and transistor currents for
MPA 1, send the following command:
MPA01<LF>
2.4.2.6.17
Query MPA Coolant Temperature and Transistor Currents
Query to get the MPA coolant temperature and transistor currents for the MPA that was initiated using the
MPA Coolant Temperature and Transistor Current initiation command.
Syntax:
MPA?
Response Format:
MPAn=t,a1,a2,a3,a4,b1,b2,b3,b4,c1,c2,c3,c4,d1,d2,d3,d4<LF>
Where: