![Aquaread BlackBox Скачать руководство пользователя страница 71](http://html.mh-extra.com/html/aquaread/blackbox/blackbox_instruction-manual_2960649071.webp)
Aquaread BlackBox™ Instruction Manual
10401-00618 Rev P
9.7. Important note concerning AP-7000 self cleaning
The MODBUS and SDI-12 protocols do not have provision for activating the AP-7000's self
cleaning mechanism.
BlackBoxs running software 2.05 and below will automatically command a cleaning cycle to
be carried out on initial power-up and each time the system is woken up from sleep mode.
In order to activate the self cleaning mechanism, toggle the on/off (sleep) line.
Remember
to allow at least 60 seconds after wake-up before sampling the data to allow the
readings to stabilise.
In order to reduce wear on the cleaning mechanism and electrodes, a maximum reading
rate of once per hour is recommend.
BlackBoxes running software revision 3.00 and above will automatically command a
cleaning cycle every six hours, regardless of the mode selected. For this reason,
power
must be left applied to the BlackBox at all times
and readings can be taken as often as
required.
If a low power mode is required between readings, use the on/off (sleep) line.
Remember
to allow at least 60 seconds after wake-up before sampling the data to allow the
readings to stabilise.
© 2019 Aquaread Ltd.
www.aquaread.com
Page 71 of 71