Aquaread BlackBox™ Instruction Manual
10401-00618 Rev P
AP-7000 continued...
Sequences initiated with Additional Concurrent Measurement command:
Command
Response
<a>C1!
or
<a>CC1!
<a>00001<CR><LF>
<a>D0!
<a><depth><CR><LF>
or
<a><depth><crc><CR><LF>
Command
Response
<a>C2!
up to
<a>C9!
or
<a>CC2!
up to
<a>CC9!
<a>00000<CR><LF>
Sequences initiated with Continuous Measurement command:
Command
Response
<a>R0!
or
<a>RC0!
<a><baro><temp><ph><orp><cond><cond20><cond25><res><sal><tds>
<ssg><CR><LF>
or
<a><baro><temp><ph><orp><cond><cond20><cond25><res><sal><tds>
<ssg><crc><CR><LF>
Command
Response
<a>R1!
or
<a>RC1!
<a><do><do-sat><aux1><aux2><aux3><aux4><aux5><aux6><nh3>
<CR><LF>
or
<a><do><do-sat><aux1><aux2><aux3><aux4><aux5><aux6><nh3><crc>
<CR><LF>
Command
Response
<a>R2!
or
<a>RC2!
<a><depth><CR><LF>
or
<a><depth><crc><CR><LF>
Command
Response
<a>R3!
to
<a>R9!
or
<a>RC3!
to
<a>RC9!
<a><CR><LF>
or
<a><crc><CR><LF>
8.4.23.
Important note concerning AP-7000 self cleaning
The SDI-12 and MODBUS protocols do not have provision for activating the AP-7000's self
cleaning mechanism. BlackBoxs running software revision 2.05 and below will
automatically command a cleaning cycle to be carried out on initial power-up and each time
the system is woken up from sleep mode.
In order to reduce wear on the cleaning
mechanism and electrodes, a maximum reading rate of once per hour is recommend.
BlackBoxes running software revision 3.00 and above will automatically command a
cleaning cycle every six hours, regardless of the mode selected. For this reason,
power
must be left applied to the BlackBox at all times
and readings can be taken as often as
required. If a low power mode is required between readings, use the on/off (sleep) line. See
section
© 2019 Aquaread Ltd.
www.aquaread.com
Page 61 of 71