![Aquaread Aquaprobe AP-5000 Скачать руководство пользователя страница 21](http://html.mh-extra.com/html/aquaread/aquaprobe-ap-5000/aquaprobe-ap-5000_instruction-manual_2960643021.webp)
Aquaprobe
®
AP-5000 Instruction Manual
10401-00906 Rev M
Select
Auto Data Logging
by pressing the down arrow key then the right arrow key or the
OK
key. The Auto Data Logging screen will be displayed.
Auto Data Logging
Interval:10 Mins
Status:OFF
Using the arrow keys to navigate, set the desired logging interval anywhere between 1 and
90 minutes.
To select permanent display logging mode, set the Status to
ON
. To select Low Power
logging mode, set the Status to
LOW POWER
.
To activate the selected logging mode, press the
OK
key then revert back to the normal
operation screen from the Main Menu by pressing the left arrow key.
To indicate that Auto Data Logging is switched on, an asterisk (*) character will flash on
and off just below the battery symbol on all the main reading screens.
If permanent display logging mode was selected (Status set to
ON
), the Meter will record a
full set of data automatically at the set rate until either the memory is full or the batteries go
flat.
If Low Power Logging Mode was selected (Status set to
LOW POWER
), the Meter will
switch itself off 30 seconds after your last key-press. Thereafter it will switch back on at the
set rate, stay on for 30 seconds, log the data, then switch back off again. This will be
repeated until either the memory is full or the batteries go flat.
If you press any key while the Meter is off between readings in low power mode, the Meter
will switch back on. If no further key is pressed, the Meter will switch back off again after 30
seconds and resume Low Power Mode.
You can cancel Auto Data Logging at any time by going back into the screen above and
setting the
Status
to
OFF
. Auto Data Logging will also be cancelled if you switch the Meter
off manually.
8.6. Battery and Memory Duration in Low Power Logging Mode
Low Power Logging Mode is specifically designed for long term data logging. In order to
estimate battery life and memory usage, the following table can be used.
The battery life figures quoted below are based on fresh, good quality alkaline batteries at
a Meter temperature of 21°C or over. Colder Meter temperatures will drastically reduce the
battery life. For example, at 5°C, the battery life will be approximately half that quoted.
© 2019 Aquaread
®
Ltd.
www.aquaread.com
Page 21 of 138