65
10.5.4
HIGH SOLUTE PASSAGE
High solute passage at normal permeate flow may have different causes.
1. Leaking O-Ring
Leaking O-rings can be detected by the probing technique.
O-rings may leak after exposure to certain chemicals or mechanical stress, e.g., element movement caused by
water hammer.
The corrective measures are:
▪
If one pressure vessel shows a significantly higher permeate concentration than the other vessels of the
same stage, then this vessel should be probed.
▪
Inspect O-rings of couplers, adapters, and end plugs for correct installation and as-new condition.
▪
Replace old and damaged O-rings.
▪
Proper shimming of the elements in a pressure vessel is essential to minimize the wear to the seals.
2. Telescoping
APRO elements can be mechanically damaged by telescoping, where the outer membrane layers of the element
unravel and extend downstream past the remaining layers. A modest telescoping does not necessarily damage the
membrane, but the glue line and/or the membrane can be ruptured in more severe cases. Telescoping is caused
by excessive pressure drop from feed to concentrate.
10.5.5
QUICK GUIDE (TABLE)
Changes of the permeate flow, the salt passage and the differential pressure are symptoms which can be attached
to specific causes in many cases. Although, the symptoms of different causes may over-lap in reality, and the
symptoms are more or less pronounced in specific cases.
Permeate
flow
Salt
passage
Differential
pressure
Direct cause Indirect cause
Corrective measure
⇑
→
Oxidation
damage
Free chlorine, ozone, KMnO
4
Replace element
⇑
→
Membrane
leak
Permeate backpressure; abrasion Replace element, improve
cartridge filtration
⇑
→
O-ring leak
Improper installation
Replace O-ring
⇑
→
Leaking
product tube
Damaged during element loading Replace element
⇓
Scaling
Insufficient scale control
Cleaning, scale control
⇓
Colloidal
fouling
Insufficient pretreatment
Cleaning, improve
pretreatment
↓
→
⇑
Biofouling
Contaminated raw water, insuffi-
cient pretreatment
Cleaning, disinfection,
improve pretreatment
⇓
→
→
Organic
fouling
Oil; cationic polyelectrolytes wa-
ter
hammer
Cleaning, improve
pretreatment
⇓
↓
→
Compaction
Water hammer
Replace element or add ele-
ments
Increasing
↓
Decreasing
→
Not changing
⇑
Main symptom
Содержание APRO 1000 LPH
Страница 1: ...aquaphor pro com TP APRO WH R02 APRO WH 1000 LPH WATER PURIFICATION SYSTEM...
Страница 8: ...8 aquaphor pro com TP APRO WH R03 2 SYSTEM OVERVIEW APRO WH 1000 LPH system dimensions...
Страница 9: ...9 aquaphor pro com TP APRO WH R03...
Страница 13: ...13 aquaphor pro com TP APRO WH R03 3 2 ELECTRIC CIRCUIT SCHEMES...
Страница 14: ...14 aquaphor pro com TP APRO WH R03...
Страница 15: ...15 aquaphor pro com TP APRO WH R03...
Страница 16: ...16 aquaphor pro com TP APRO WH R03...
Страница 17: ...17 aquaphor pro com TP APRO WH R03...
Страница 18: ...18 aquaphor pro com TP APRO WH R03...
Страница 19: ...19 aquaphor pro com TP APRO WH R03...
Страница 20: ...20 aquaphor pro com TP APRO WH R03...
Страница 80: ...80 14 ADDITIONAL MATERIALS...
Страница 81: ...81...
Страница 83: ...83 REMARKS...