START UP PROCEDURE
1/
Install STAC2 vertically taking in account instrument constraint and recommendation
2/
Make the electrical connections required following the required functions (MODBUS, Ethernet, USB)
3
/
Open the front door of the flowing case.
4/
Install and Connect the FlowCell cuvette to the Pump output tube and STAC2 main output tubes.
(
Warning :
FlowCell cuvette is delivered in its own package outside the STAC2 case).
WARNING :
Each FlowCell cuvette has its own FlowCell size Factor written on the
bottom of FlowCell cuvette (sticker). This value MUST be entered in the STAC2
software parameter to give you correct measurement results (see FlowCell size
factor description in software description chapter)
Flow-through
cuvette