STAC2
- 52 -
9.7.2 RINSE SETTINGS
The next area, is dedicated to the Rinse channel. Rinse fluid solution is a specific fluid usefull to clean
internal STAC2 tubing and flow cell windows inside a certain limit.
The customer can decide if this function is enabled or not, the frequency of this cleaning process and if
the cleaning occur at begenning or end of a measurement cycle.
Generally, it’s recommanded to clean system “after sample measurement” to clean potential dust
deposite any Sample has dispersed in tubing and flow cell.
Depending on the target application and chemical analitic paramaters, a diluted javel water solution is a
good cleaning sample. This remove and clean algaes and others particle.
Set the frequency measurement. This is a
multiple of main cycle process measurement
defined in general configuration.
Ex:
10min main cycle and frequency of 60 result in a
Rinsing process every 60 min = 1 hour.
Total tube length
for rinse channel
in meter.
Pump sucking
duration will be
automatically
adjusted
according tube
length outside
the STAC2.
Consequently,
Rinse solution
automatically
reach the flow
cell correctly.
Select, when
rinsing action
occur inside a
cycle
measurement
process.
Enabling or
disabling Rinse
function.
Содержание Secomam STAC2
Страница 1: ...USER MANUAL version 1 03 ON LINE ANALYZER ...
Страница 2: ...STAC2 2 INSTALLATION USER S MANUAL ...
Страница 5: ...STAC2 5 ...
Страница 7: ...STAC2 7 ...
Страница 11: ...STAC2 11 ...
Страница 13: ...STAC2 13 ...
Страница 14: ......
Страница 27: ......
Страница 31: ......
Страница 59: ...STAC2 59 ...
Страница 65: ...STAC2 65 ...
Страница 74: ......