6.10
PROGRAM Mode Operation
Program mode applies to AC output mode and allows output settings and expected load
measurement limits to be defined using up to nine sequential steps. Measurement data is
taken at each step and compared to preset user defined high and or low limits. If one of
more measurements is outside the limits, a FAIL is generated and the test sequence is
aborted. If no limit violations are detected during this program execution, a PASS result is
generated instead.
6.10.1
PROGRAM Mode System Settings
To select PROGRAM mode of operation, make sure the output is OFF first. Changing mode
is not possible while the output is active.
From the top level Menu, select the “
System
” soft key. This brings up the System menu. The
selected operating mode will be shown in the top left field as the first parameter entry.
In PROGRAM mode, the following system parameters can be set from this screen:
•
Auto Run Mode
•
Single Step Mode
•
Alarm Volume
•
Contrast of LCD Display
•
Power Up Mode
•
Loop Cycle Count
•
Results display Mode
•
Over Current Protection Fold Back mode
•
Lock Mode
•
Memory Lock Mode