6RL SERIES OPERATION MANUAL
SECTION 3: PRODUCT OVERVIEW
Entire Contents Copyright
2018 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
6RL Series Regenerative DC Load Operation Manual
Page 24 of 204
3.8.3
Analog Interface
This 15 pole Sub-D socket on the rear panel of the unit is provided for remote
control of the unit via analog or digital signals. If remote control is in
operation this analog interface can only be used alternately to the digital
interface. Also, monitoring is always available.
The input voltage range of the set values and the output voltage range of the
monitor values, as well as reference voltage levels can be configured in the
SETTINGS menu of the unit between as 0-5 V or 0-10 V, in each case for 0-
100%.
3.8.4
Share Connections
The two-pole WAGO style socket (labelled “Share”) on the rear
panel of the unit is provided for connection to same sockets on
compatible APS power supplies series to achieve a balanced load
current distribution during parallel connection. The same socket
is also used to connect the power supply to compatible APS electronic loads, in order to
build a two-quadrant operation setup. Following power supply and electronic load series are
compatible:
APS DCS Series
APS 6RL Series
3.8.5
Voltage Sense Connections (Remote Sense)
In order to compensate for voltage drops along the DC cables
from the connected DC supply, the Sense input can be connected
to the source being tested. The maximum possible voltage drop
compensation is provided in the technical specifications.
Note: Remote sensing is only intended for constant voltage (CV) operation and it is
recommended to have the “Sense” input only connected to the source
under test when running the load in CV mode. In other regulation modes,
but also in CV, the sense cables can cause unwanted side effects, like
oscillation, based on their length and inductance.
Note: In order to ensure safety and to comply with international directives,
insulation of high voltage models, i.e. such with a nominal voltage of 500 V
or higher, must be ensured by using only the two outer pins of the 4-pole
terminal. The inner two pins, marked with NC, must remain unconnected at
all times.