APPENDIX C
AUTO SEQUENCE PROGRAMMING
Entire Contents Copyright
2014 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
44M04 LOAD MAINFRAME OPERATIONS MANUAL
Page 86 of 93
44M00
10.
Repeat steps 6 to 9 to as needed to enter up to 16 steps to form the auto-sequence.
11.
Once the desired number of steps have been set, press the STORE button.
12.
The LCD will show REP (repetitions).
13.
Use the arrow keys to set the number of auto-sequence repetitions.
14.
Press STORE to confirm the sequence edit.
This completes the programming sequence.
8.3
Test Mode
To execute a previously stored auto-test sequence, proceed as follows:
1.
Press the TEST key on the mainframe to enter the TEST mode.
2.
Use the numbered STATE keys (1 to 9) to select the previously saved auto-sequence.
3.
Press ENTER to start the auto-sequence.
4.
The LCD shows “PASS” or “FAIL” after testing.
If limits and the NG functions have been set and a test step fails, the mainframe LCD
display will flash “NG”. The user must then press ENTER to continue the auto-
sequence execution or EXIT to abort the auto-sequence.
5.
Press Auto-sequence or EXIT at any time to abort an auto-test sequence.