CHAPTER 4
REMOTE CONTROL PROGRAMMING
Entire Contents Copyright
2014 by Adaptive Power Systems, Inc. (APS) • All Rights Reserved • No reproduction without written authorization from APS.
44M04 LOAD MAINFRAME OPERATIONS MANUAL
Page 69 of 93
44M00
4.6.3
STATE Commands
STATE commands can be used to set or query the actual operating status of the electronic
load at any time.
[STATe:] LOAD {SP} {ON | OFF}
Command Syntax:
[STATe:] LOAD {SP} {ON | OFF}{; | NL}
[STATe:] LOAD? {; | NL}
Purpose:
Read LOAD ON or OFF status.
Description:
This command is used to see if the Load is ON or OFF. 0 = Load OFF1, = Load ON.
[STATe:] MODE {SP} {CC | CR | CV | CP | LED}
Command Syntax:
[STATe:] MODE {SP} {CC | CR | CV | CP | LED} {; | NL}
[STATe:] MODE? {; | NL}
Purpose:
Set and read the operating mode of LOAD
Description:
The return value is 0 │ 1 │ 2 │ 3 │ 4 which corresponds to the operating mode that the
load is in. i.e. CC│CR│CV│CP │LED.
Mode:
CC
CR
CV
CP
LED
Value:
(0)
(1)
(2)
(3)
(4)
41Lxxxx
V
V
V
V
42Lxxx
V
V
V
V
41Dxxx
V
V
V
V
V
42D5003
V
V
V
V
[STATe:] SHORt {SP} {ON | OFF}
Command Syntax:
[STATe:] SHORt {SP} {ON | OFF}{; | NL}
[STATe:] SHORt? {; | NL}
Purpose:
Reads back whether the short circuit test is active or not
Description:
0 = short circuit test active, 1 = short circuit test inactive