44M04 LOAD MAINFRAME OPERATIONS MANUAL
Page 47 of 93
44M00
AUTO SEQUENCE COMMANDS
NOTES
RETURN
T2 {SP} {NR2} {; | NL}
0.1~9.9(s)
0.1~9.9(sec)
SAVE {; | NL}
Save “File n” data
REPEAT {SP} {n} {; |
NL}
n=0~9999
0~9999
RUN {SP} {F} {n} {; | NL}
n=1~9
AUTO REPLY
“PASS” or “FAIL:XX”
(XX=NG STEP)
Table 13: Short Form – Auto Sequence Commands
4.4.1
Notations and Conventions Used in programming commands:
1.
Frequency engineering unit: Hz
2.
GLOB:
GLOBAL (all channels active at the same time).
3.
Current engineering unit:
A
4.
Voltage engineering unit:
V
5.
Resistance engineering unit: Ω
6.
Time Period engineering unit: ms
7.
Slew-rate engineering unit:
A/us or mA/us
8.
Power engineering unit:
W
9.
*9 MPPT and CR DYN function supported from FW version as follows:
a)
44M04 at r2.15 version and higher enabled.
b)
41L0630 at r1.11 version and higher enabled.
c)
41L0660 at r1.11 version and higher enabled.
d)
41L2512 at r1.11 version and higher enabled.
e)
41L5012 at r1.11 version and higher enabled.
f)
41L0615 at r1.11 version and higher enabled.