2.9
Interface Options
The 44M04 support one of four different remote control interface options. The interface
specified at the time of order is installed at the factory prior to shipment. It is possible to
retrofit interface options in the field. Contact Adaptive Power Systems Customer Service for
instructions.
2.9.1
RS232 Serial Interface
Figure 3 shows the RS232 connector (Female) on the rear panel. This connects the 44M04
mainframe to RS232 port of computer. The RS232 BAUD-RATE can be set on the front
panel of the 44M04. Press the “SYSTEM” button twice to enter the desired BAUD RATE
adjustment.
Figure 3:44M04 RS232 Connection
2.9.2
GPIB Interface
The GPIB connector is located on the rear panel. This socket allows the 44M04 to be
connected to the controller and other GPIB devices. A GPIB system can be connected in any
configuration (star, linear, or both) as long as the following conditions are met:
•
The maximum number of devices including the controller is equal or less than 15.
•
The maximum length of the GPIB cable is no more than 2 meters.
•
The total lead length of all devices connected together totals less than 20 meters.
•
Please make sure the lock screws are firmly hand-tightened, use a screwdriver only for the
removal of screws. Figure 4 shows the rear panel of 44M04 mainframe. The GPIB address
of the 44M04 mainframe is set on the front panel.
Figure 4: 44M04 GPIB Connection