Document 4207Q258 V3.00 January 2018
Page 19 of 30
Chirascan Couette Cell LD Accessory User Manual
Figure 3.3: Interior of the sample chamber with the accessory in place
3.2.2 Temperature control connections
Although the main components of the accessory are temperature rated from 5°C to 95°C,
some of the standard fittings for the circulating water are only rated to 60°C, and could fail at
higher temperatures, causing injury to the user and damage to the accessory. The standard
operating range should be taken as 5°C to 60°C. Do not use the accessory at temperatures
outside this range unless the standard fittings have been replaced with high temperature
alternatives by Applied Photophysics.
There is no direct temperature control on the Couette Cell Accessory. However, the temperature can be
controlled by an external a water circulator with a flow rate of about 200
– 300 ml per minute. To achieve this,
a pressure of approximately 0.2
– 0.3 bar (3 to 4.5 p.s.i.) is needed.
If the water pressure of the circulator used for controlling the temperature is too high, the tubing
can be forced from the connectors within the spectrometer sample housing unit. Do not exceed
an input water pressure of 1.7 bar (24 p.s.i.).
The circulating water connections are to the male Legris type fittings on the tubing at the rear of the sample
chamber (Figure