You can use the QuantStudio
™
6 or 7 Flex System touchscreen to view instrument
version data, such as instrument serial number and version numbers for most
instrument components and firmware.
1.
From the Main Menu of the instrument touchscreen, touch Tools, then touch
About the Instrument.
2.
When you are finished, touch to return to the Main Menu.
The QuantStudio
™
6 and 7 Flex Systems feature a secure mode that can be enabled to
restrict local instrument functionality. When security is enabled, use of the
touchscreen is restricted to administrative functions that change the instrument
settings. After the system is secured, you must enter an administrator password to
modify the instrument settings, use the firmware tools, or deactivate the secure mode.
IMPORTANT!
If you enable or disable the QuantStudio
™
6 and 7 Flex Systems
security, auditing, and electronic signature feature, you must similarly enable or
disable the QuantStudio
™
Real-Time PCR Software security (see “Access the Security
screen and enable or disable security“ on page 131). The QuantStudio
™
Real-Time
PCR Software cannot connect to the QuantStudio
™
6 and 7 Flex Systems that do not
match security settings.
Note:
Secure mode limits the number of features that are available from the system
touchscreen. It does not provide user authentication functionality through the
instrument touchscreen.
1.
If the instrument is in standby, tap the touchscreen to activate it, then tap .
2.
In the main menu, tap Settings, then tap Set Administrator Options.
3.
In the Set Administrator Options screen, tap Secure Environment to enable
(checked) or disable (unchecked) system security.
4.
(Optional) To change the administrator password:
a.
Tap Change Password.
b.
Enter the current password, then tap Done.
c.
Enter the new password, then tap Done.
d.
Reenter the password when prompted.
e.
Tap OK when prompted.
Note:
The default password for the instrument touchscreen is password;
however, the password can be changed during installation.
5.
Tap Save.
6.
Tap the Administrator Password field, enter the administrator password, then
tap Done.
7.
Tap to return to the main menu.
Review the
instrument
version
information
Enable or disable
instrument
security
Appendix A
Manual instrument operation
Administrate the instrument from the touchscreen
A
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide
95