8.
Seal the array card(s):
a.
With the carriage (roller
assembly) of the TaqMan
®
Array
Card Sealer in the Start position,
place a filled array card into the
fixture with the foil side up so that
the fill reservoirs are the farthest
away from the carriage.
b.
Press down on all four corners of
the array card to ensure that it is
fully seated within the fixture.
c.
Use the two alignment pins in the
fixture to position the array card
correctly.
d.
Seal the array card by running the
carriage slowly over it. Run the
carriage over the array card in one
direction only. Do not apply
downward force on the carriage
as you move it forward over the
card.
e.
Remove the sealed array card
from the fixture and trim the fill
reservoirs from the array card
assembly using scissors. Trim the
foil array card so that the edge is
even with the plastic carrier.
IMPORTANT!
Completely
remove the fill reservoirs from the
array card so that the edge is free
of residual plastic. The plastic
from the fill reservoirs that
extends beyond the edge of the
card can prevent the array card
from seating properly on the
sample block and can affect
amplification.
Chapter 2
Calibrate the instrument
Preparing array cards for instrument calibration
2
40
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide