• Safety glasses
• Powder-free gloves
• Plate adapter
1.
Touch the instrument touchscreen to activate it, then press .
2.
From the Main Menu, touch .
3.
When the tray arm opens, pull the latch, then lift and remove the plate adapter.
4.
Attach the new adapter to the tray arm, then pull the latch to allow the adapter to
lower into place. If necessary, apply pressure as indicated until the adapter snaps
into place.
5.
From the Main Menu, touch .
6.
If you are installing a tray adapter of a different consumable format, replace the
sample block and heated cover if necessary.
Materials required
Change the plate
adapter
Appendix B
Maintenance
Change the plate adapter
B
112
QuantStudio
™
6 and 7 Flex Real-Time PCR Systems v1.6 Maintenance and Administration Guide