![Applied Biosystems Data Explorer 4 Series Скачать руководство пользователя страница 327](http://html.mh-extra.com/html/applied-biosystems/data-explorer-4-series/data-explorer-4-series_user-manual_2955117327.webp)
Voyager Data Examples
Data Explorer
™
Software User’s Guide
7-11
7
7.2 Voyager Data Examples
This section includes:
•
Detecting and labeling partially resolved peaks
•
Processing before calibrating to optimize mass accuracy
•
Detecting peaks from complex digests
7.2.1 Detecting and Labeling
Partially Resolved Peaks
If peaks are not
labeled
If peaks are partially resolved and the peaks of interest are not
labeled, you can adjust the following peak detection
parameters:
Figure 7-9 shows a spectrum with partially resolved peaks that
represent two compounds.
If...
Adjust the following...
Peaks represent two
compounds, and you
want both peaks labeled
In the Peak Label dialog box, select Allow Overlapping
Peak Labels.
OR
On the Basic Settings tab in Peak Detection, set %Max
Peak Area to 0, then adjust the %Base Peak Intensity
until peaks are detected.
OR
On the Peak Processing tab in Peak Detection, change
the default Integration Baseline Setting from
Valley-to-Valley to Valley-to-Baseline.
Peaks represent partially
resolved isotopes, and
you want to label and
detect the average mass
On the Basic Settings tab in Peak Detection, decrease
the Mass Resolution setting until the isotopic envelope is
detected.
Содержание Data Explorer 4 Series
Страница 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Страница 10: ...Table of Contents x Applied Biosystems ...
Страница 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Страница 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Страница 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Страница 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Страница 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Страница 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Страница 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Страница 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Страница 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...
Страница 446: ...I D E X N Index 30 Applied Biosystems ...