Chapter 5 Examining Spectrum Data
5-44
Applied Biosystems
5
4.
Click OK.
The trace is displayed with an RSM, NF, NR or SM
trace label.
5.
To return to the original trace, see “Returning to the
original spectrum” on page 5-3.
High-resolution data
Noise
Removal
(NR)
(Does not
affect peak
resolution.)
Specify the number of standard deviations of
noise to remove. The software automatically
calculates the average white noise for all
frequencies across the spectrum, then
removes the specified number of standard
deviations of noise.
This method slightly affects peak intensity, and
removes peaks with a signal-to-noise ratio less
than the specified standard deviation.
In general:
If you set
% White Noise
Std Dev to:
removed is:
1
68
2
95
3
99
If applying Noise Removal with a certain Std
Dev does not yield the necessary noise
removal, return to the original trace (see
“Returning to the original spectrum” on
page 5-3) and apply Noise Removal again with
a higher Std Dev setting. Applying Noise
Removal more than one time with the same
Std Dev setting does not improve noise
removal.
Type of Data
Suggested
Method
Description
Содержание Data Explorer 4 Series
Страница 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Страница 10: ...Table of Contents x Applied Biosystems ...
Страница 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Страница 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Страница 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Страница 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Страница 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Страница 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Страница 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Страница 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Страница 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...
Страница 446: ...I D E X N Index 30 Applied Biosystems ...