Chapter 3 Peak Detection and Labeling
3-60
Applied Biosystems
3
Charge state
not displayed
If no charge is displayed, there are a few possible causes:
•
Peaks are more than 1 Da apart.
•
Filter width is set too high to detect other isotope peaks.
•
The maximum charge state for charge state
determination is set lower than the charge state of the
peak.
•
Charge state determination parameters are set such that
peaks are determined to have no charge. See
Section 3.2.5, Charge State Determination and
Examples.
•
Peaks are not from the same isotope species.
NOTE: If you set %Base Peak Intensity or %Max Peak
Area too low (described on page 3-20), noise peaks
can be detected, and charge states can be incorrectly
labeled.
Содержание Data Explorer 4 Series
Страница 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Страница 10: ...Table of Contents x Applied Biosystems ...
Страница 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Страница 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Страница 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Страница 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Страница 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Страница 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Страница 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Страница 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Страница 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...
Страница 446: ...I D E X N Index 30 Applied Biosystems ...