Peak Detection
Data Explorer
™
Software User’s Guide
3-31
3
Filter Settings
Width
Number of data points used in smoothing for peak detection
before integration. This value is automatically calculated by
the software if you select Use Resolution Dependent Settings
on the Basic Settings tab. For more information, see
Section 3.1.2, The Resolution-Based Peak Detection
Routine.
NOTE: If you set Filter Width too high, narrow peaks may not
be detected.
Hint: If you manually set Filter Width, set to a number equal
to the number of points across the peak. To determine the
number of points across a peak, you can change the trace
display from Line to Vertical Bars. Each vertical bar
represents one data point. For more information, see
Section 1.4, Customizing the Data Explorer Window.
Increment
Number of data points the filter moves across. Use 1 for all
applications.
This value is automatically set to 1 if you enable Use
Resolution Dependent Settings on the Basic Settings tab.
For more information, see Section 3.1.2, The Resolution-
Based Peak Detection Routine.
Table 3-4 Advanced Settings (Spectrum Data Only) (Continued)
Parameter
Description
Содержание Data Explorer 4 Series
Страница 1: ...Data Explorer Software Version 4 Series Software User Guide ...
Страница 10: ...Table of Contents x Applied Biosystems ...
Страница 56: ...Chapter 1 Data Explorer Basics 1 42 Applied Biosystems 1 ...
Страница 98: ...Chapter 2 Using Chromatogram and Spectrum Windows 2 42 Applied Biosystems 2 ...
Страница 172: ...Chapter 3 Peak Detection and Labeling 3 74 Applied Biosystems 3 ...
Страница 204: ...Chapter 4 Examining Chromatogram Data 4 32 Applied Biosystems 4 ...
Страница 270: ...Chapter 5 Examining Spectrum Data 5 66 Applied Biosystems 5 ...
Страница 316: ...Chapter 6 Using Tools and Applications 6 46 Applied Biosystems 6 ...
Страница 384: ...Chapter 9 Troubleshooting 9 22 Applied Biosystems 9 ...
Страница 388: ...Appendix A Warranty A 4 Applied Biosystems A ...
Страница 416: ...Appendix C Data Explorer Toolbox Visual Basic Macros C 20 Applied Biosystems C ...
Страница 446: ...I D E X N Index 30 Applied Biosystems ...